Advertisement

4.Gas charge
4.1 Evacuation②
Preparations: Dedicated GM, Refrigerant Cylinder, Bottle Cap, Powerful Vacuum Pump, Scale
(Electronic Weighing Scale), Charge hose, branching tool, tool
Procedure : After operating the powerful vacuum pump with the refrigerant cylinder on the scale,
Open the valve in the following order and evacuate to the refrigerant cylinder.
VS⇒VR⇒VG⇒VC.
Vacuuming time: about 1 minute with a vacuum pump with capacity of 300 L / min
(Standard: vacuum degree 3 Torr or less)
Main valve
Scale

4.2 Completion of evacuation

Procedure : Specified time (about 1 minute) after evacuating,
shut the valve in the following order and stop the pump.
VC⇒VG⇒VR⇒VS
Remove the vacuum pump from the hose.
Dedicated GM
Refrigerant
VR
port
VS
VC
Vacuum
port
Charge hose
Powerful vacuum pump
-46-
VH
VG
VL
Valve state
open
shut
VC
VG
VL
VH
VR
VS
Main
SVL
Valve state
VC
VG
VL
VH
VR
VS
main
SVL
High pressure
port
Low pressure
port
SVL
Low pressure side
service valve
open
shut