Setting Up Generic and Advanced Signals
Instrument Settings
Channel Settings
38
Setting Up a Pulse Signal
Task
Set up a continuous pulse signal with 20 ns period, a pulse width
of 5 ns, an amplitude of 2.0 V and an offset of 1.5 V (high level 2.5
V, low level 500 mV).
Using the Graphical User Interface
To set the instrument parameters:
1 Disable the outputs.
2 In the instrument panel, choose the Pulse/Pattern mode.
3 Set the Period to 20 ns.
To set the channel parameters:
1 Switch to the Channel page.
2 Select the pattern mode Pulse.
3 In the Timing panel:
– Switch from DCyle to Width and set the Width to 5 ns.
4 In the Levels panel, set:
– The level format to Custom.
– Switch from High/Low to Amplitude/Offset.
– Amplitude to 2.0 V.
– Offset to 1.5 V.
Keysight 81133A/81134A Pulse Generator User's Guide, April 2015
Setting Up a Pulse Signal