Summary of Contents for Keysight Technologies Y1165A

  • Page 1 Keysight Y1165A Kit Removable SSD for the M9384B Installation Guide...
  • Page 2 WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS documentation. No additional DOCUMENT THAT CONFLICT WITH government requirements © Keysight Technologies, Inc. THESE TERMS, THE WARRANTY beyond those set forth in the 2020 TERMS IN THE SEPARATE EULA shall apply, except to the AGREEMENT WILL CONTROL.
  • Page 3 Where to Find the Latest Information Documentation is updated periodically. For the latest information about this product, including instrument software upgrades, application information, and product information, browse to the following URL: https://www.keysight.com/find/m9384b To receive the latest updates by email, subscribe to Keysight Email Updates at the following URL: http://www.keysight.com/find/MyKeysight Information on preventing instrument damage can be found at: http://keysight.com/find/PreventingInstrumentRepair...
  • Page 5: Table Of Contents

    Locations for Keysight Technologies ........
  • Page 6 Y1165A Installaton Guide...
  • Page 7: Installing The Y1165A Ssd

    M9384B VXG Microwave Signal Generator Initialization Guide Installing the Y1165A SSD The following topics can be found in this section: “Overview” on page 8 “SSD Interconnect Life Expectancy” on page 9 “SSD Removal and Installation” on page 10 “Firmware Updates” on page 12...
  • Page 8: Overview

    Installing the Y1165A SSD Overview Overview The Keysight Y1165A provides an additional removable Solid State Drive (SSD) for your M9384B VXG. The SSD is mounted in a receptacle on the top of the M9384B chassis and can be easily inserted or removed. This is convenient when multiple people are using the same VXG and they do not want to interfere with the other person’s...
  • Page 9: Ssd Interconnect Life Expectancy

    Installing the Y1165A SSD SSD Interconnect Life Expectancy SSD Interconnect Life Expectancy The life expectancy of both the connector on the removable SSD and the mating connector internal to the instrument is 1,500 cycles. If this is exceeded it is very likely that a failure will occur between these connections. Keysight highly recommends that this cycle limit not be exceeded to prevent an unwanted failure of the instrument.
  • Page 10: Ssd Removal And Installation

    Installing the Y1165A SSD SSD Removal and Installation SSD Removal and Installation Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a static-safe workstation. Refer to the documentation that pertains to your instrument for information about static-safe workstations and ordering static-safe accessories.
  • Page 11 Installing the Y1165A SSD SSD Removal and Installation 4. Gently pull the SSD connector away from the mating connector on the M9384B top cover. 5. Install the new SSD by sliding the SSD connector into the mating connector on the M9384B.
  • Page 12: Firmware Updates

    Installing the Y1165A SSD Firmware Updates Firmware Updates To determine the currently installed instrument firmware version on the VXG, select the Gear icon > System > Firmware Version. If an instrument firmware update is required, the latest revision of the firmware can be found by going to the product page: https://www.keysight.com/find/m9384b...
  • Page 13: Instrument Security Information

    Installing the Y1165A SSD Instrument Security Information Instrument Security Information Information on instrument security features and the instrument volatility can be found at: http://www.keysight.com/find/security Y1165A Installaton Guide...
  • Page 14 Installing the Y1165A SSD Instrument Security Information Y1165A Installaton Guide...
  • Page 15: Initializing The Y1165A Ssd Accessory For Use In The M9384B

    M9384B VXG Microwave Signal Generator Initialization Guide Initializing the Y1165A SSD Accessory for Use in the M9384B The Keysight Y1165A provides an additional removable Solid State Drive (SSD) for your M9384B VXG. The following procedures will accomplish the initialization process: “Setting the Instrument’s Serial Number”...
  • Page 16: Requirements

    Initializing the Y1165A SSD Accessory for Use in the M9384B Requirements Requirements — PathWave signal generator version A.3.02, or higher, running on the SSD — Mouse and keyboard connected to the instrument Y1165A Installaton Guide...
  • Page 17: Setting The Instrument's Serial Number

    Initializing the Y1165A SSD Accessory for Use in the M9384B Setting the Instrument’s Serial Number Setting the Instrument’s Serial Number 1. From the Windows Start menu, select > Utilities > Interactive IO. 2. From the Connect menu, select Connect. 3. In the Resource Name text box, type TCPIP0::localhost::hislip1, then select OK.
  • Page 18 Initializing the Y1165A SSD Accessory for Use in the M9384B Setting the Instrument’s Serial Number 4. In the Command field type :SERV:PROC "VXG Microwave Signal Generator,M9384B,US12345678", where US12345678 is the serial number of your instrument. Select Send Command. 5. Verify the setting by typing *IDN? into the Command area then select Send &...
  • Page 19: Establishing The Instrument's Hardware Configuration File

    Initializing the Y1165A SSD Accessory for Use in the M9384B Establishing the Instrument’s Hardware Configuration File Establishing the Instrument’s Hardware Configuration File 1. Select the System Menu , > Exit then select OK to exit the M9384B application. 2. From the Windows Start menu, select File Explorer and navigate to E:\Keysight\PathWave\SignalGenerator You will find at least five file names sghal<_Fxx_xxx>.cfg.
  • Page 20 Initializing the Y1165A SSD Accessory for Use in the M9384B Establishing the Instrument’s Hardware Configuration File If there was already a sghal.cfg file in the folder, overwrite it with the one that matches your instrument. Do not rename the original file as you may need it if you reuse this SSD in another M9384B.
  • Page 21: Setting The Instrument's Computer Name

    Initializing the Y1165A SSD Accessory for Use in the M9384B Setting the Instrument’s Computer Name Setting the Instrument’s Computer Name 1. This procedure must be performed in the Administrator account. To sign out of the default user (Instrument) account, go to the Windows Start menu and select the Change Account settings icon and sign out.
  • Page 22 Initializing the Y1165A SSD Accessory for Use in the M9384B Setting the Instrument’s Computer Name 6. In the System Properties dialog, select Change. 7. In the Computer Name/Domain Changes dialog, enter your desired Computer Name. (The default factory Computer name is K-M9384B-12345, where 12345 is the last 5 digits of the instrument’s...
  • Page 23 Initializing the Y1165A SSD Accessory for Use in the M9384B Setting the Instrument’s Computer Name 8. Select OK. You will now be prompted to restart the instrument. Make sure to save any open files and close all programs, then select OK to continue.
  • Page 24: Verify Configuration

    Initializing the Y1165A SSD Accessory for Use in the M9384B Verify Configuration Verify Configuration 1. If upon restart the bootup screen shows a Hardware Failure message similar to the one below, the incorrect sghal.cfg file was copied during “Establishing the Instrument’s Hardware Configuration File.
  • Page 25 Initializing the Y1165A SSD Accessory for Use in the M9384B Verify Configuration 3. In the System Information window, verify the Serial Number and Computer Name are as you expect. — If either of these are incorrect, repeat “Establishing the Instrument’s Hardware Configuration File.
  • Page 26 Initializing the Y1165A SSD Accessory for Use in the M9384B Verify Configuration Y1165A Installaton Guide...
  • Page 27: Safety And Maintenance Information

    M9384B VXG Microwave Signal Generator Initialization Guide Safety and Maintenance Information The following topics can be found in this section: “Safety Information” on page 28 “Warnings, Cautions, and Notes” on page 29 “Instrument Markings” on page 32 “Instrument Maintenance” on page 33 “Returning an Instrument for Service”...
  • Page 28: Safety Information

    10°C from the lowest, of the maximum operating temperature of a single instrument. If there are any concerns or special requirements a Keysight Field Engineer should be consulted to assure instrument(s) temperature compliance and performance. Y1165A Installaton Guide...
  • Page 29: Warnings, Cautions, And Notes

    Do not proceed beyond a caution until you fully understand the indicated conditions. Note calls the user’s attention to an important point or special information in the text. Y1165A Installaton Guide...
  • Page 30: General Safety Considerations For The M9384B

    The measuring terminals on this instrument are designed to be used with external signals described in Measurement Category I, but NOT with external signals described in Categories II, III, and IV. The input of this instrument cannot be connected to the mains. Y1165A Installaton Guide...
  • Page 31: Servicing

    These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so. Operating Conditions This product is designed for use in Installation Category II and Pollution Degree 2 environment. Y1165A Installaton Guide...
  • Page 32: Instrument Markings

    Universal recycling symbol. This symbol indicates compliance with the China standard GB 18455-2001 as required by the China RoHS regulations for paper/fiberboard packaging. The Keysight email address is required by EU directives applicable to our product. Y1165A Installaton Guide...
  • Page 33: Instrument Maintenance

    Wash thoroughly after handling. In case of spill, soak up with sand or earth. Flush spill area with water. Dispose of isopropyl alcohol in accordance with all applicable federal, state, and local environmental regulations. Y1165A Installaton Guide...
  • Page 34: Returning An Instrument For Service

    Returning an Instrument for Service Returning an Instrument for Service Calling Keysight Technologies Keysight Technologies has offices around the world to provide you with complete support for your instrument. To obtain servicing information or to order replacement parts, contact the nearest Keysight Technologies office listed below.
  • Page 35: Service Options

    Safety and Maintenance Information Returning an Instrument for Service Service Options Keysight Technologies offers several optional maintenance plans to service your instrument after the warranty has expired. Call your Keysight Technologies office for full details. If you want to service the instrument yourself after the warranty expires, you can download the service documentation that provides all necessary troubleshooting and maintenance information from the Keysight web page.
  • Page 36 This information is subject to change without notice. © Keysight Technologies 2020 Edition 1, April 2020 Y1165-90001 www.keysight.com...