EMC Notice

Electromagnetic emissions

The HC271/HC271p monitors are intended for use in the electromagnetic environment specified below. The
customers or the users of the HC271/HC271p monitors should assure that they are used in such an
environment.
Table 1-3
Electromagnetic emissions
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000–3–2
Voltage fluctuations/ flicker emissions
IEC 61000–3–3

Electromagnetic immunity

Table 1-4
Electromagnetic immunity
Immunity test
Electrostatic discharge (ESD)
IEC 61000–4–2
Electrical fast transient/burst
IEC 61000–4–4
Surge
IEC 61000–4–5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000–4–11
6
Chapter 1 Getting Started
Compliance
Group 1
Class B
Class D
Complies
IEC 60601–1–2:2014 Test
levels
±8 kV contact
±15 kV air
±2 kV for power supply lines
±1 kV line(s) to line(s)
±2 kV line(s) to earth
< 5% U
1 (> 95% dip in U
) for
T
T
0.5 cycle
40% U
(60% dip in U
) for 5
T
T
cycles
70% U
(30% dip in U
) for 25
T
T
cycles
< 5% U
1 (> 95% dip in U
) for
T
T
250 cycles
Electromagnetic environment - Guidance
The HC271/HC271p uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
The HC271/HC271p is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings
used for domestic purposes.
Compliance level
±8 kV contact
±15 kV air
±2 kV for power supply lines
±1 kV line(s) to line(s)
±2 kV line(s) to earth
< 5% U
1 (> 95% dip in U
) for
T
T
0.5 cycle
40% U
(60% dip in U
) for 5
T
T
cycles
70% U
(30% dip in U
) for 25
T
T
cycles
< 5% U
1 (> 95% dip in U
) for
T
T
250 cycles
Electromagnetic environment
— Guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
The main power quality should
be that of a typical commercial
or hospital environment.
The main power quality should
be that of a typical commercial
or hospital environment.
The main power quality should
be that of a typical commercial
or hospital environment. If the
user of the HC271/HC271p
requires continued operation
during interruptions in the main
power, it is recommended that
the HC271/HC271p be powered
from an uninterruptible power
supply or a battery.