Installing the GC
Step 2 Verify line voltage, voltage settings, and
power cord
1 Locate the power label near the power cord connector on the back of the GC. Compare the
2 Verify that the power cord is correct for the voltage and location.
Electrical shock hazard. To avoid injury, only a qualified person should measure line voltage.
WAR N IN G
3 Have a qualified person measure the actual power outlet voltage and verify that it meets
The following sections detail the power specifications and requirements for reference.

Line voltage

Installation and First Startup
instrument power settings with the laboratory line voltage. See
page 16.
available"
on page 17.
the tolerance requirements listed in
"Line voltage"
on page 15.
Typical GC System - 8890 GC with computer and printer.
Leave 12-in. (31 cm)
open space for
operational access
Table 1
on page 16. See
Use isolated circuits with
dedicated grounds for GC and
31-in. (79 cm)
(without heat
deflector)
26-in. (66 cm)
(with heat
deflector)
"Power consumption"
"Power cords
"Grounding"
on page
21"
on
and
15