Grounding; Line Voltage; Line Power Cord Terminations - Agilent Technologies 7890B Installation And Startup

Gas chromatograph
Hide thumbs Also See for 7890B:
Table of Contents

Advertisement

Grounding

Line voltage

Line power cord terminations

Agilent 7890B Installation
To protect users, the metal instrument panels and cabinet are
grounded through the three-conductor power line cord in
accordance with International Electrotechnical Commission
(IEC) requirements.
The three-conductor power line cord, when plugged into a
properly grounded receptacle, grounds the instrument and
minimizes shock hazard. A properly grounded receptacle is one
that is connected to a suitable earth ground. Proper receptacle
grounding should be verified.
Make sure the GC is connected to a dedicated receptacle
(outlet).
The GC operates from one of the AC voltage supplies listed in
Table
1, depending on the standard voltage of the country from
which it was ordered.
GCs are designed to work with a specific voltage; make sure
your GC voltage option is appropriate for your laboratory. The
voltage requirements for your GC are printed near the power
cord connector.
Although your GC should arrive ready for operation in your
country, compare its voltage requirements with those listed in
Table
1. If the voltage option you ordered is not suitable for your
installation, contact Agilent Technologies.
The power cord termination is determined by the country
where the GC is ordered.
1