Calibrate Air Flow - Bosch ACS 561 Repair Instructions

A/c service unit
Table of Contents

Advertisement

Diagnostics and Testing
6. If Continue (F1) is pressed, the following
displays:
7. If the calibration is not accepted, a Calibra-
tion Failed screen is displayed. If accepted,
the accepted screen displays:
Failed:
8. Pressing Exit (F2) returns to Calibration
Load Cells in the Service Menu.
If the check repeatedly fails, check that
the weights are placed on the scale
properly.
SP00D00517
2018-07-11
Accepted:

Calibrate Air Flow

The air flow inside the unit must be present at
the correct volume to make sure enough air is
present to prevent the buildup of refrigerant and
to keep the compressor temperature below its
upper limit.
WARNING: During airflow calibration, all the
unit plastic enclosures must e onto the unit
and properly attached. This is because the
enclosure affects the calibration itself. The unit
then must be in standard operating conditions.
1. Scroll to Calibrate Air Flow and press Select
(F1). The following is displayed:
2. Press Calibrate (F1). The following displays:
22
ACS 561 Service Manual
Robert Bosch GmbH