Bioland 2006-2B User Manual page 13

Table of Contents

Advertisement

Power frequency
(50Hz/60Hz)
30 A/m
magnetic field
IEC61000-4-8
Guidance and manufacture's declaration – electromagnetic immunity
The 2006-2B Blood Pressure Monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of the device should assure
that it is used in such an environment.
IEC 60601
Compliance
Immunity test
test level
level
Conducted
3 Vrms
3 Vrms
RF
150 kHz ~
150 kHz to
IEC61000-4-6
80 MHz
80 MHz
6Vrmsc)in
6Vrmsc)in
ISM bands
ISM bands
between
between
0,15 MHz
0,15 MHz
and
80
and
MHz
MHz
Radiated RF
IEC61000-4-3
10 V/m
10V/m
80 MHz to
Power
frequency
magnetic fields should
be
at
30 A/m
characteristic
typical location in a
typical commercial or
hospital environment.
Electromagnetic
environment-guidance
Portable
and
mobile
communications equipment should be
used no closer to any part of the
2006-2B Blood Pressure Monitor,
including
cables,
than
recommended
separation
calculated
from
the
equation
applicable to the frequency of the
transmitter.
Recommended separation distance:
80
150 kHz to 80 MHz
d
P
1
2 .
80 MHz to 800 MHz
d
1
2 .
P
800 MHz to 2.7 GHz
d
2
3 .
P
d=6
/E
at
RF
communications
equipment
(Portable
RF
communications
levels
of
a
RF
the
distance
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
a. Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
wireless
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
bands
To assess the electromagnetic environment due to fixed RF transmitters, an
Page 13 of 14
80 MHz ~
2.7 GHz
equipment (including peripherals such
2.7 GHz
as
antenna
antennas) should be used no closer
than 30 cm (12 inches) to any part of
the 2006-2B Blood Pressure Monitor,
including cables specified by the
manufacturer).
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer
recommended separation distance in
metres (m).
Field
transmitters, as determined by an
electromagnetic site survey, a should
be less than the compliance level in
each frequency range. b
Interference may occur in the vicinity
of
equipment
following symbol:
cables
and
external
and
d
is
the
strengths
from
fixed
RF
marked
with
the

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Bioland 2006-2B

Table of Contents