Blood Pressure Monitor
Page 1 of 14
Model: 2006-2B
User manual

Advertisement

Table of Contents
loading

Summary of Contents for Bioland 2006-2B

  • Page 1 Blood Pressure Monitor Model: 2006-2B Bioland Technology LTD. No. A6B7 (Block G) ShangRong Industrial Zone No. 5 Baolong Road, Baolong Community Longgang District 518116, Shenzhen, Guangdong PEOPLE’S REPUBLIC OF CHINA Toll-free service line: 800-218-2838 Mon-Fri 9AM-5PM EST Date of issue: 2019.04.10...
  • Page 2: Table Of Contents

    Table of Content Introduction Table of Content......................2 Introduction........................2 Thank you for purchasing this blood pressure monitor. This remarkable instrument Precaution before use.....................3 is ideal for people who frequently monitor their own blood pressure because it's Few words about blood pressure................... 4 easy to use.
  • Page 3: Precaution Before Use

    11. Any reading may be affected by the measuring position, the patient's position, Precaution before use movement, or the patient's physical condition 12. Failure to use the cuff specified by use manual, or modified instrument may Do not confuse self-monitoring with self-diagnoses. Blood pressure result in incorrect measurements.
  • Page 4: Few Words About Blood Pressure

    24. There is no allergic reaction to the accessible material on the device; Please consult your physician to determine if your blood pressure accurately reflects 25. Children are not allowed to operate the device and should not eat any falling your actual blood pressure parts;...
  • Page 5: Product Description

    * Talking * Taking a bath * Drinking alcohol Note: There is not an universally accepted definition of hypotension (low blood * Smoking, etc. pressure), but a systolic pressure below 99 mmHg is usually regarded as Note: One or two readings will not provide a true indication of your normal blood hypotension pressure.
  • Page 6: Install Or Replace Batteries

    Environment Protection – Waste electrical products should not disposed of with household waste. Please recycle where facilities exist. Check with your local Install or replace batteries authority or retailer for recycling advice. When the battery symbol appears on the display or nothing is display while inflation How to use the upper arm cuff during measurement, you should replace with new batteries.
  • Page 7: A Few Suggestion Before Measurement

    better cross reference; Important! The mark (about 2 cm long bar) must lie exactly over the artery which 2. When measuring, the legs should not be crossed, feet flat on the ground, arms runs down the inner side of the arm. and back vertical.
  • Page 8: How To Measure Blood Pressure

    How to measure blood pressure Remain still and do not talk during the measurement. Position your arm higher than the heart will get lower measure result, contrarily the measure result will get higher. 1. Set up the cuff to your upper arm as previous section of "How to use the upper This unit can intelligently adjust the cuff pressure and inflate to a higher pressure level when needed, so its normal when the unit re-inflate during one measurement.
  • Page 9: Recalling Memory

    Recalling memory Maintenance and repair Press “M” button to turn on the device, then press “M” button until enter 1. Do not wash or immerse the unit in water, use a soft and memory state like right figure. dry cloth to clean it. 2.
  • Page 10: Specifications

    Specifications Symbols information Indication Digital LCD display Pressure display Range: Pressure: 0-300 mmHg, Pulse rate: 40~195 Beat/min Type BF applied part Pressure display Accuracy: Pressure: ±3mmHg; Pulse rate: ±5% Attention and read before use SYS 60 to 260 mmHg DIA 30 to 195 mmHg NIBP measurement range Indicates the medical device manufacturer Pulse rate 40 to 195 /min...
  • Page 11: Troubleshooting

    Cuff not properly adjust or E0, E2, E7, E8, E9 Renew adjust cuff position and placement, hand or body The 2006-2B Blood Pressure Monitor is intended for use in an electromagnetic Measurement failure tightness, and keep hand still shaking. environment in which radiated RF disturbances are controlled. The customer or Re-measurement or send back E3.
  • Page 12 IEC 61000-4-2 15 kV air relative humidity should The 2006-2B Blood Pressure Monitor is intended for use in the electromagnetic be at least 30%. environment specified below. The customer or the user of the device should assure ±2kV for power that it is used in such an environment.
  • Page 13 Guidance and manufacture’s declaration – electromagnetic immunity watts (W) according to the transmitter The 2006-2B Blood Pressure Monitor is intended for use in the electromagnetic manufacturer environment specified below. The customer or the user of the device should assure recommended separation distance in that it is used in such an environment.
  • Page 14: Fcc Requirement

    If the measured field strength in installed and used in accordance with the instructions, may cause harmful the location in which the 2006-2B Blood Pressure Monitor is used exceeds the interference to radio communications. However, there is no guarantee that applicable RF compliance level above, the SD1Ultrasonic Doppler should be interference will not occur in a particular installation.