manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
Guidance and manufacturer's declaration – electromagnetic emission
The 2006-2B Blood Pressure Monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of the device should assure
that it is used in such an environment.
Emission test
Compliance
RF emissions
Group 1
CISPR 11
RF emission
Class B
CISPR 11
Harmonic emissions
Not applicable
IEC/EN61000-3-2
Voltage fluctuations
/flicker emissions
Not applicable
IEC/EN61000-3-3
Guidance and manufacture's declaration–electromagnetic immunity
The 2006-2B Blood Pressure Monitor is intended for use in the electromagnetic
Electromagnetic environment -
guidance
The
2006-2B
Blood
Pressure
Monitor uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
The
2006-2B
Blood
Pressure
Monitor is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings
used
for
domestic
purposes.
environment specified below. The customer or the user of the device should assure
that it is used in such an environment.
IEC 60601 test
Immunity test
level
Electrostatic
discharge (ESD)
IEC 61000-4-2
±2kV for power
Electrical Fast
Transient/Burst
IEC/EN61000-4-4
input/output lines
± 1 kV line(s) to
Surge
IEC/EN61000-4-5
± 2 kV line(s) to
<5%UT(>95% dip
Voltage dips, short
40%UT(60%dip
interruptions, and
voltage variations
on power supply
70%UT(30%dip
input lines
IEC/EN61000-4-11
<5%UT(>95% dip
Page 12 of 14
Compliance
Electromagnetic
level
environment-guidance
Floors should be wood,
concrete or ceramic tile.
8
kV
8 kV contact
If floor are covered with
contact
15 kV air
synthetic material, the
15 kV air
relative humidity should
be at least 30%.
supply lines
Not
Not applicable
±1kV for
applicable
line(s)
Not
Not applicable
applicable
earth
in UT)
for 0.5cycle
in UT)
for 5 cycles
Not
Not applicable
applicable
in UT)
for 25 cycles
in UT)
for 5s