Proline Promass F 300 EtherNet/IP
Diagnostic information
No.
383 Memory content
Status signal
Diagnostic behavior
Diagnostic information
No.
387 HistoROM backup failed
Status signal
Diagnostic behavior
Diagnostic information
No.
303 I/O 1 to n configuration changed
Status signal
Diagnostic behavior
Diagnostic information
No.
330 Flash file invalid
Status signal
Diagnostic behavior
Diagnostic information
No.
331 Firmware update failed
Status signal
Diagnostic behavior
Diagnostic information
No.
410 Data transfer
Status signal
Diagnostic behavior
Short text
F
Alarm
Short text
F
Alarm
12.8.3

Diagnostic of configuration

Short text
M
Warning
Short text
M
Warning
Short text
F
Warning
Short text
F
Alarm
Diagnostics and troubleshooting
Remedy instructions
1. Restart device
2. Delete T-DAT via ' R eset device'
parameter
3. Replace T-DAT
Remedy instructions
Contact service organization
Remedy instructions
1. Apply I/O module configuration
(parameter ' A pply I/O
configuration' )
2. Afterwards reload device
description and check wiring
Remedy instructions
1. Update firmware of device
2. Restart device
Remedy instructions
1. Update firmware of device
2. Restart device
Remedy instructions
1. Check connection
2. Retry data transfer
Coding of diagnostic
information (hex)
0x100016E
Coding of diagnostic
information (hex)
0x1000288
Coding of diagnostic
information (hex)
0x400026C
Coding of diagnostic
information (hex)
0x40002C9
Coding of diagnostic
information (hex)
0x10002CA
Coding of diagnostic
information (hex)
0x100008B
201