Security Code Page - ABB 8235 Chloride Instruction Manual

Ion-selective electrode monitor
Table of Contents

Advertisement

...6 PROGRAMMING
...6.2.2 Operating Page 2
Continued from previous page.
Calibrating Std 2
Electrode mV
CALIBRATION PASSED
or
CAL FAILED (SLOPE)
or
CAL FAILED (SLOW)
or
Mode
Chloride

6.2.3 Security Code Page

A security code is required to gain access to all subsequent programming pages. The code is preset at the factory to '0' but can be
changed if required in the Current Output Page – see Section 6.2.5.
Security Code -----
or
Incorrect code.
SET UP INPUT
Chloride
14
Calibrating Standard 2
The upper display shows the chloride concentration value based on the previous good
calibration figures. The lower display remains until a stable output is obtained from the
electrodes.
Toggle between the two lower displays.
– –
Electrode Millivolts
The electrode output can also be displayed during calibration
Cal
normal operation.
When a stable output is detected the display automatically advances to the next
parameter, and the upper display is updated to the value of "Std 2".
Calibration Passed
A satisfactory calibration has been carried out.
or
Calibration Failed (slope)
The monitor was unable to obtain an acceptable slope value during calibration. The
slope value required is 80 to 110%.
or
Calibration Failed (slow)
The monitor was unable to obtain a stable output from the electrodes during
calibration.
Return to top of Operating Page 2.
or
Return to Operating Page 1.
<Unit>
Security Code
Set the correct security code.
Advance to the first of the programming pages – Set Up Input Page.
<Units>
Return to Operating Page 1.
Note. Pressing
during a calibration aborts the sequence and returns to
Cal