...6 PROGRAMMING
6.2.1 Operating Page 1
The values displayed in Operating Page 1 are for viewing only and cannot be altered in this page.
Chloride
<Unit>
or
Millivolts
or
Control Temp
x x
or
Slope
x x x
or
Date
DD: MM:YY
or
Time
HH : MM: SS
or
Next AutCal
DD: MM:YY
or
Last Cal
DD: MM:YY
or
or
SECURITY CODE -----
12
Measurement Units
The measurement units are displayed: Chloride mg l
Mode
Advance to next parameter.
x x
xmV
Sensor Output
The electrode output is displayed in millivolts.
Mode
Advance to next parameter.
°C
x
Control Temperature
The measured flowcell temperature is displayed in degrees Celsius.
Mode
Advance to next parameter.
%
x
Slope Check Value
The slope value should be between 80 and 110%. If the value is outside these
limits check the electrode.
Advance to next parameter.
Mode
Date
The date is displayed.
Mode
Advance to next parameter.
Time
The time is displayed.
Mode
Advance to next parameter.
Next Calibration Date
The date when the next automatic calibration is to be carried out is displayed. If the
automatic calibration is disabled, 'OFF' is displayed in place of the date.
Mode
Advance to next parameter.
Last Calibration Date
The date of the last automatic or manual calibration is displayed.
Mode
Return to top of Operating Page.
or
Advance to Security Page (Section 6.2.3).