...6.2.11 Electrical Calibration Page
Continued from previous page.
Adjust Output 1
or
Enter
Adjust Output 1
or
Enter
Adjust Output 2
or
Enter
Adjust Output 2
or
Enter
26
Zero
Adjust Current Output 1 Zero
The monitor transmits a zero signal, e.g. for a 4 to 20 mA output range, 4 mA is
transmitted.
Set the milliammeter reading to the current output 1 zero level, i.e. 0 mA (zero-based
ranges) or 4 mA (4 to 20 mA range).
Store.
Advance to next parameter.
FSD
Adjust Current Output 1 Full Scale
The monitor transmits a span signal, e.g. for a 4 to 20 mA output range, 20 mA is
transmitted.
Set the milliammeter reading to the current output 1 full scale level, i.e. 1 mA, 10 mA or
20 mA as applicable.
Store.
Advance to next parameter.
zero
Adjust Current Output 2 Zero
The monitor transmits a zero signal, e.g. for a 4 to 20 mA output range, 4 mA is
transmitted.
Set the milliammeter reading to the current output 2 zero level, i.e. 0 mA (zero-based
ranges) or 4 mA (4 to 20 mA range).
Store.
Advance to next parameter.
FSD
Adjust Current Output 2 Full Scale
The monitor transmits a span signal, e.g. for a 4 to 20 mA output range, 20 mA is
transmitted.
Set the milliammeter reading to the current output 2 full scale level, i.e. 1 mA, 10 mA or
20 mA as applicable.
Store.
Advance to next parameter.