Download Print this page

Philips PM8926/59 Instruction Manual

10:1 passive probe

Advertisement

Quick Links

10:1 PASSIVE PROBE
PM8926/59
Instruction manual
4822 872 03343
880909
1. INTRODUCTION
PHILIPS
This probe is a 1 Ox attenuator probe with a total length of 150 em approx., designed for
oscilloscopes.
The probe is provided with a special BNC plug with built-in resistor for automatic range
indication to advances the V/DIV reading of the oscilloscope by lOx.
NOTE: This provision is only possible in oscilloscopes equipped with probe
indication feature!
At delivery the probe has been adjusted to an oscilloscope with an input
capacitance of 20
pF.
2. CHARACTERISTICS
-
Attenuation (d.c.)
-
Useful system bandwidth
-
Input resistance at d.c.
-
Input capacitance at d. c.
-
Compensation range
-
Max. non destructive input
voltage (d.c. + a.c. peak)
at d. c.
3. ACCESSORIES
\......-
- Accessory kit, contents:
-
Instruction manual
4. DISMANTLING
lOx +or-2%
d.c
....
lOO MHz (-3 dB)
10 Mn
+or-
1%
14 pF +or- 1.5 pF
15 pF ... 35 pF
500
v
- Ground lead
- Retractable hook tip
- DIL-IC tip
- Coloured marking tags (2)
- Insulating cap
- Ground bayonet
To expose the h.
I.
adjustment control the terminating box cover can be slid off the
compensation box.