Troubleshooting

Intstallation
JuLI Stage does not
power up
LCD screen is black
Monitoring
Poor image
(Too dark and too bright)
Dirty image
Time-lapse images become
dark or bright
Problem for saving movie
file
Check on/off switch on back side of main device.
Check power source or contact your distributor.
Move the mouse.
Check if power supply is connected and power switch is on.
Reset the power button.
Re-optimize the brightness and exposure value.
Reset the power button
Swipe the stage with cotton swab carefully.
Eliminate any dust on the culture dish.
Check for and remove any condensation on the lid of the culture dish.
Warm up the instrument for 2 hours before monitoring. (power should
be on when warming up the instrument)
Cell culture flask should be wetted by the culture media before making
a movie.
Eliminate any dust on the culture dish.
Remove any condensation on the lid of the culture dish.
To prevent any problem such as shaking of culture dish or inflowing light,
pay special attention when you open or close incubator door during
monitoring.
Check available volume of the hard disk.