e.g.) For the pressure range 0 to 10 kPa (0 to 100% F. S.), change the product displayed
with 0.00 to 10.00 kPa, to 0 to 1020.
Decimal point position
Display value of the minimum pressure range
Display value of the maximum pressure range
7.2.5 Selecting the filter
The product is equipped with a digital filter with 5 different time constants. Use it when
pressure fluctuation is intense preventing easy reading.
 This time constant also reflects the comparator and analog output of
the selected filter.
Use
key to select "Select Filter". After
current setting is displayed. Use
changed.
7.2.6 Scaling of analog output (with output options)
This mode sets the pressure corresponding to the analog output zero point (4mA DC or 1V
DC) / span point (20mA DC or 5V DC).
Use
key to change to "Analog Scaling". After
pressure corresponding to the current analog output zero point (4 mA DC or 1V DC) is
displayed as a percentage of the full scale of the pressure range. Enter the value which is
changed by
keys.
Use
key and
as well.
e.g.) Change the product with 4 to 20mA DC output in the range of 0 to 10 kPa (0 to
100% F. S.) to the same output in the range of 0 to 9.81kPa.
Pressure at the analog output zero point: 0.0%F.S.
(4mA DC output in the pressure rage of 0%F.S.)
Pressure at the analog output span point: 100%F.S.
(20mA DC output in the pressure rage of 98.1%F.S.)
CAUTION
keys to select time constant of the filter to be
‐‐‐‐‐‐ No filter
‐‐‐‐‐‐ Time constant 25ms
‐‐‐‐‐‐ Time constant 250ms
‐‐‐‐‐‐ Time constant 2.5s
‐‐‐‐‐‐ Time constant 5s
‐‐‐‐‐‐ Time constant 10s
keys to set the pressure of the analog output span point (
17
2 →
:
0 →
:
1000 →
:
is displayed for one second, the
is displayed for one second, the
TY-GC62-010A Rev.2
0
0.0
1020
)
0.0%F.S.