 When setting value is manipulated in the Function Setting Mode, all
setting values, including comparator, are recalculated and setting
value is overwritten. If the recalculated setting value falls outside
the display range, the setting value will be the maximum value or
the minimum value that can be internally processed.
7.2.2 Selecting comparator operation
Use
key to select "Select Comparator Operation". After
second, the current setting appears. Use
comparator.
7.2.3 Selecting the display
Use
key to select "Select Display". After
current setting appears. Use
kPa) or display scaling.
Pressure range
0 to 50 Pa
0 to 100 Pa
0 to 200 Pa
0 to 0.5 kPa
0 to 1 kPa
0 to 2 kPa
0 to 5 kPa
0 to 10 kPa
0 to 20 kPa
* In case of compounding, (-) is used for negative pressure display.
7.2.4 LED display scaling
If you select
in "Select Display", the values to be displayed on LED display is
scaled to an arbitrary value against the applied pressure. This function scales the LED
display value to the minimum/maximum value of the pressure range, and does not affect
the relationship between applied pressure and analog output.
Use
key to change to "Display Scaling". After
decimal point in the current Measurement Mode appears. Use
decimal point.
Use
and
keys to set the display value of minimum pressure range (
and display value of maximum pressure range (
Since the set value of the display scaling is stored internally, when display goes back to SI
units display in "Select Display", you can use it by selecting
resetting.
CAUTION
keys to select either hysteresis or window
keys to select either pressure display in SI units (Pa or
Max. differential
pressure value
Compound
Pa
+/- 50 Pa
50.0
+/- 100 Pa
100.0
200.0
+/- 200 Pa
200
+/- 0.5 kPa
-
+/- 1 kPa
-
-
+/- 2 kPa
-
+/- 5 kPa
-
+/- 10 kPa
-
-
<Table 7-1>
16
is displayed for one
is displayed for one second, the
Display scaling
kPa
etc.
-
-
-
-
0.500
1.000
2.000
2.00
5.00
10.00
20.00
is displayed for one second, the
keys to change the
) as well.
again without
TY-GC62-010A Rev.2
),