Table of Contents

Advertisement

6. CONNECTING

CONNECTING | BASE + XY TABLE (1/3)
4.
3.
Connect USB 2.0 (see options below) (1+2), control box (3),
monitor (4), ethernet (5), USB 3.0 (6), ESD (Only black models) (7)
and power supply (8) in that order.
USB 2.0 (1+2): Keyboard, mouse, USB key, USB hard drive, scanner
Connect keyboard and mouse in the USB 2.0 inputs (only for use
with some applications). Keyboard and mouse need to be HID
compliant.
You can also use the USB inputs for a USB key, USB hard drive
and/or barcode scanner. If necessary use a USB hub to get
multiple USB ports.
Connector Description
1+2
USB 2.0 input
3
Control box
4
Output
5
Ethernet
6
USB 3.0 output
7
ESD
8
Power supply
(DC)
6.
5.
7.
Type
Connect to
USB 2.0 Type A Socket
Keyboard
Mouse
USB key
USB hard drive
Barcode scanner
D-SUB 9-Pin Female
Control box
HDMI Standard Type A
Monitor
RJ-45
Ethernet wall socket
USB 3.0 Type B Socket
Computer
Snap Socket, 10mm
ESD Grounding point
DC Jack
Power socket
Live image
1.
2.
8.
The microscope can be connected directly to a monitor.
Live image and image capturing through USB3.0 / 3rd part software.
See system recommended (third-party) under system requirements.
You are still able to connect monitor(s) as described above.
Specification
USB 2.0 Host. Output: 5V
Output: 3.3V
Use only control boxes supplied by TAGARNO
HDMI Out 1080p60
Ethernet LAN, 100BASE-TX/1000BASE-T
USB 3.0 Device. Self-powered
ESD Grounding point. Only black models
Input: 12V
Use only the power supply provided by TAGARNO:
Brand/Model: Mean Well/GSM40A12
Protection Class I
Input: 100-240V
Overvoltage category ll
Output: 12V
Live image
/ image capture
USB3.0
HDMI - HDMI cable
500mA
6A Max
1.3A
1.0-0.5A, 50/60Hz
3.34A Max
6