11. Do not dismantle any parts of the microscope, except where noted in the manual 12. Never disassemble or clean internal optical surfaces 13. Use only the power supply provided from TAGARNO 14. Always turn off the system before unplugging, when possible Warning: This is a Class A product.
5. yOu HaVE RECEiVED Manual Camera arm rod, washers, screws usB Memory stick Hexagon tool power supply HdMi - dVi Cable usB 3.0 a-B Cable OpTiOns Xplus fHd control box HdMi - HdMi cable XKey Control box Base Xy table...
6. assEmBlinG assEmBlinG | BasE (1/2) How to install video: tagarno.com/video-tutorials/how-to-install-tagarno-trend.mp4 assemble base and rod with the screw. tighten the rod both below pull the button and place the camera arm on the rod the base vertically and horizontally on top of the base...
Page 5
assEmBlinG | xy TaBlE (2/2) assemble Xy table, block and rod with three screws pull the button and place the camera arm on the rod fasten the two washers with the screw on top of the rod the complete microscope with Xy table...
Connect to Specification power supply dC Jack power socket input: 12V 1.7a (dC) Use only the power supply provided by tAGARno: Brand/Model: Mean well/gsM40a12 input: 100-240V 1.0-0.5a, 50-60Hz output: 12V 3.34a Max Control box d-suB 9-pin female Control box output: 3.3V...
Page 7
COnnECTinG | EsD (2/3) Connect esd grounding cord (electrostatic discharge) Connect esd grounding cord (electrostatic discharge) only black models only black models...
Page 8
DVI out DVI out The microscope an be connected to multiple monitors by adding The microscope can be connected directly to a monitor an active signal splitter between the TAGARNO and the monitors Live image / image capture USB3.0 Live image and image capturing through USB3.0 / 3rd part software.
8. OpERaTiOn OpERaTiOn | BasE + xy TaBlE (1/3) How to use video: tagarno.com/video-tutorials/how-to-use-tagarno-trend.mp4 place the camera arm in the appropriate height according to the adjust the friction tension to set your personal preference lens you are using power on/off 1.
Page 10
NB! Never look directly into the laser pointer. Setup Menu If you are using another TAGARNO unit than TAGARNO FHD Access the Setup menu by pressing the round center TREND, a window on the screen will provide you with information.
Page 11
Factory default directs to the Ruler application with both horizontal og vertical ruler activated. Multi Function Buttons XKEY Control box for TAGARNO FHD functions. When the Setup menu is deactivated, you are able to use Do only connect equipment distributed by TAGARNO.
9. sETup mEnu sETup mEnu (1/4) The setup menu in your TAGARNO FHD microscope provides you NB! If the camera’s field of view does not contain a white object, with 2 entrances to setup your microscope exactly as you like it.
Page 13
sETup mEnu (2/4) 2. general settings 7. System units Choose to display metrics or imperial units when using the microscope. 8. Numeric format Here you can select the numeric format based on your preference. Choose between 1,234.56, 1.234,56 or 1 234,56. 9.
Page 14
sETup mEnu (3/4) 3. appliCations When the Ruler is chosen the system will automatically start in manual focus set to focus on the part of the object farthest away. It is also possible to save an image with graphics from the Ruler function.
Page 15
(4/4) 4. inFo 5. Verification lines TAGARNO measurement application is required in order to run this application. The application provides a range of options to control the lines: Recall calibrations from TMA to use as starting point for placing...
10. mEasuREmEnT appliCaTiOn (Tma) tagarno MeasureMent appliCation | Calibration mEasuREmEnT appliCaTiOn | CaliBRaTiOn (1/3) To use the measurement application the mouse and keyboard must be Click Run Calibration and check if the result says PASS connected as described on page 6.
Page 17
MeasureMent appliCation | Calibration anD tool Menu mEasuREmEnT appliCaTiOn | CaliBRaTiOn anD TOOl mEnu (2/3) saving anD reCalling a Calibration new Calibration - CustoM DotgriD Advanced calibration is done automatically by placing any The application will store and recall a calibration, so that you do not have predefined dot grid calibration tool under the camera to recalibrate every time the software has been closed.
Page 18
MeasureMent appliCation | tool Menu mEasuREmEnT appliCaTiOn | TOOl mEnu (3/3) Background color adjustment Adjust the color of the text annotation background Text Annotation before using the text annotation tool. The hue is set to This function makes it possible to add text annotations transparent by default.
This is a step by step guide to perform a system update on your TAGARNO FHD TREND, PRESTIGE or UNO. Plug the USB memory stick into one of the two USB 2.0 inputs at the back of the microscope. Leave the other USB 2.0 input empty. • To install a new application or updates to your TAGARNO, you need to perform a system update on your current microscope. • To perform a system update, a system update file needs to be USB 2.0 input installed directly on the microscope.
Page 20
Please see the user manual for further information regarding the use of TAGARNO applications. If you have trouble performing a system update please contact the TAGARNO distributor from which you have purchased the product or send an e-mail to support@tagarno.com. When the system update is complete, the microscope will automatically restart.
12. liCEnsE aCTiVaTiOn A. Access the Setup menu by pressing the round center After installation of a TAGARNO application, you might need to activate button on the XPLUS FHD control box until the Setup menu a unique license key file to be able to access and use the application.
Backlight kit glass table pOwER supply Description power supply type dC Jack Input microscope input: 12V 1.7a Warning use only the power supply provided by tagarno Brand/Model Mean well/gsM40a12 Input power supply 100-240V 1.0-0.5a, 50-60Hz output 3.34a Max...
14. sysTEm REquiREmEnTs FULL HD 1080p @ 50/60Hz Desktop Laptop Windows 7, 8 or 10 (with DirectX 11) Memory Intel® Core™ i5 or i7 @2.4GHz, CPU: Intel® Core™ i5 or i7 @2.4GHz (3th or 4th generation named 3xxx or (4th generation named 4xxx) 4xxx) Hard Disc Space 1GB required (SSD type recommended)
The warranty only applies if the product is packed, shipped, stored, handled and maintained correctly. Failures due to improper packing and transport are not covered. Failures due to drops and sudden blows are not covered. Failures due to storage and handling in extreme high or low temperatures are not covered. Failures due to storage or handling in extreme high humidity are not covered. 17. liCEnsE aGREEmEnT this tagarno product contains open source and third-party components. tagarno.com/license-agreement ergonoMical recoMMendations 18. ERGOnOmiC RECOmmEnDaTiOns Please be careful when carrying the product from one place to another. Always lift the microscope with one hand around the rod and the other hand beneath the base. Please do not lift the microscope by grabbing the camera arm. All TAGARNO products are marked with a product label: Values marked with X varies from product to product and are replaced by specific numbers and letters.
PHONE +45 76251111 DESCRIPTION TAGARNO A/S hereby declares that the product listed above, consisting of a camera unit and a 12V power supply, is in compliance with the following European directive: EMC DIRECTIVE 2014/30/EU By conforming to the following harmonised standards:...