Table of Contents

Advertisement

3. Pull the fuse holder out of the instrument.
4. Remove the fuse from the holder and discard it into an appropriate waste container.
Figure 6-2. Replacing the Fuse
CAUTION
Use only the specified fuse to avoid damage to the instrument.
5. Replace the defective fuse with an identical fuse, snapping it into place. Fuse
specifications are in Section 1: Introduction, while ordering information is in Section
8: Service, Supplies, and Replacement Parts.
A spare fuse, located in the small, enclosed compartment in the fuse holder, is shipped
with the instrument. Push the fuse out of the compartment using a small screwdriver.
6. With the flat side up, return the fuse holder to its position in the instrument.
7. Firmly press the fuse cover until it snaps into place and is flush with the power module
plate.

Replacement of Light Assemblies

Use the following procedure to replace the Power Light Assembly (PN 94000787) or Low
Stain Light Assembly (PN 94001073).
1. Turn the instrument off and remove the line cord from the AC electrical outlet.
2. Raise the lid of the instrument.
Page 6—4, Section 6: Minor Replacements and Adjustments