2.2 Electrical Information

2.2 Electrical Information
The Compact 2A has several connectors located on the rear of the unit. A connection diagram is
provided for convenience.
Label Description
1
Thermocouple Sensor Input
2
AC Input
3
Analog Output
4
Thermocouple Sensor Ground

2.3 Calibration

The Compact 2A controller provides a calibration adjustment to maximize the accuracy of each sen-
sor. The adjustment should be set at 1 mTorr. Maximum accuracy will utilize a 2A sensor and a
NIST traceable reference gauge co-located on a vacuum chamber. A controlled leak of dry nitrogen
is necessary for calibration. All Compact 2A controllers purchased with a 2A sensor will be factory
calibrated and there is no need for adjustment.
For calibration, pump the vacuum chamber down to below 1
on the vacuum system, adjust the pressure in the chamber until the reference gauge reads 1 mTorr.
Then adjust the potentiometer located in the rear of the unit until the display shows 1 mTorr.
email: sales@frederickscompany.com
Figure 2: Compact 2A Electrical Connection Information
tel: +1 215 947 2500
4
10
Torr. Using the leak control
web: frederickscompany.com
2 SETUP
comp_2a_im rev -
5 of 7