Welch Allyn Propaq LT Directions For Use Manual page 71

Vital signs monitor
Hide thumbs Also See for Propaq LT:
Table of Contents

Advertisement

Directions for Use
Figure 74. NIBP: reading in progress
STEWART, ANN
STEWART, ANN
NIBP in progress.
NIBP in progress.
II
II
HR/min
HR/min
NIBP
NIBP
NIBP Mode:
NIBP Mode:
145
145
When the cycle completes, control returns to the previous direct-access display. To
view the final NIBP readings on the manometer scale, select Manometer from the
Setup NIBP menu.
Figure 75. NIBP: reading complete
STEWART, ANN
STEWART, ANN
II
II
PR/min
PR/min
NIBP
NIBP
NIBP Mode:
NIBP Mode:
2 2
Note
8. If motion artifact such as shivering or coughing interferes with NIBP readings, do the
following:
7762940
7762940
Adult
Adult
Rm 263
Rm 263
1mV/cm
1mV/cm
NIBP mmHg
NIBP mmHg
Resp/min
Resp/min
Manometer
Manometer
Systolic
Systolic
Diastolic
Diastolic
Manual
Manual
50 50
100
100
150
150
200
200
Exit
Exit
Trends
Trends
Snapshots
Snapshots
7762940
7762940
Adult
Adult
1mV/cm
1mV/cm
79
79
132/83
132/83
NIBP mmHg
NIBP mmHg
(98)
(98)
Resp/min
Resp/min
Manometer
Manometer
Systolic
Systolic
Diastolic
Diastolic
Manual
Manual
50 50
100
100
150
150
200
200
Exit
Exit
Trends
Trends
Snapshots
Snapshots
If you need to stop the reading and vent the cuff at any time during the
measurement, press
If the monitor cannot get a valid NIBP reading, it displays the message
'NIBP retry in progress' in the upper left corner of the screen. Then, depending on
settings and patient mode, the monitor attempts one or two more
measurements. If the retries fail to produce a valid reading, the monitor displays
an NIBP error message.
For every inch the cuff is placed above the heart, add 1.9 mmHg (0.253 kPa) to the
displayed NIBP reading. For every inch below the heart, subtract 1.9 mmHg
(0.253 kPa).
Position the patient's limb away from the body so that the applied cuff is not in
contact with the patient's body or any other object (such as a bed rail). Keep the
cuff as close to heart level as possible.
Verify that the Smartcuf filter is ON. (See
Smartcuf"
on page 68.)
Verify that the ECG leads are properly connected to the patient, and monitor ECG
during NIBP. (ECG monitoring is required for Smartcuf.)
3:00:06P
3:00:06P
% %
SpO2
SpO2
Mean
Mean
Manometer pressure indicator bar
250
250
300
300
Setup
Setup
3:00:06P
3:00:06P
Rm 263
Rm 263
Systolic, diastolic, and MAP
(numeric)
% %
SpO2
SpO2
Mean
Mean
Systolic, diastolic, and MAP
(graphical)
250
250
300
300
Setup
Setup
.
"Improving NIBP accuracy with
Standalone monitoring
67