Utility:remote; Utility:remote:mode - National Instruments RMX-400 Series Programming Manual

Programmable dc electronic load
Table of Contents

Advertisement

Query Example

:UTILity:REMote

43B
Description
Syntax
Parameter
Example

:UTILity:REMote:MODE

44B
Description
Syntax
Parameter
Example
RMX-4000 Series Programming Manual
0
1
:UTIL:SOUN?
0
Turns the remote control on or off.
:UTILity:REMote {OFF|0|ON|1}
OFF/0
ON/1
:UTIL:REM 1
Sets the remote mode to fast or normal. In fast
mode, the panel interface is deactivated with an
interface time of no more than 10 ms. Normal
mode has an interface time of 30~130 ms. In
normal mode, the display interface continues to
update the screen in real time.
:UTILity:REMote:MODE {NORMAL|0|FAST|1}
NORMAL/0
FAST/1
:UTIL:REM:MODE 1
Off
On
Sound is currently set to
off.
System Command
Turns Remote control off
Turns remote control on
Turns remote control on.
System Command
NORMAL
FAST
Turns remote mode to
fast.