    Testing sample(s) ......................18 CANNON CANNON CANNON CANNON CANNON High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual ® CANNON CANNON CANNON Revision 1.0d—June, 2011; CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 3 SPARE PARTS LIST 10 WARRANTY/RETURN INFORMATION Products limited warranty....................29 Reagent and chemical warranty ..................29 Returning a product to CANNON® ................. 30 APPENDIX A—HTHS CAPILLARY REPLACEMENT Replacing an HTHS capillary ................... 31 Parts/Tools required: ..................... 31 Removing the viscometric cell ................31 Removing the capillary ..................
  • Page 4: Introduction

    The HTHS Series II is capable of testing at a temperature of 150°C as specified in ASTM D 5481. For other applications, temperature may be set as low as 25°C above ambient.
  • Page 5: Hths Specifications

    (DEC99) requires HTHS viscosities from 2.6 to 3.7 mPa•s. Because the HT75 standard has a nominal viscosity of 2.7 mPa•s and the HT150 ® standard has a nominal viscosity of of 3.7 mPa•s, CANNON recom- mends that the users calculate cell volumes using either of these stan- dards.
  • Page 6: Operator Safety

    Operator safety Please observe the following safety procedures and notices for proper operation of your HTHS Series II instrument. Deviation from the installa- tion, operation or maintenance procedures described in this manual may result in a hazardous situation and may void the manufacturer's warranty.
  • Page 7 Refer all service and repairs to qualified personnel. • Do not attempt to service the unit beyond the service and/or repair procedures detailed in this manual or other documentation provided by the manufacturer. Contact CANNON ® Instrument Company for all additional service/repair needs.
  • Page 8: Apparatus Description

    The container slides into place at the left front of the HTHS Series II. A hinged door provides protection from hot oil. At the rear, an internal fan draws vapors into the exhaust manifold.
  • Page 9: Operation

    HTHS internal rear view including pressure relief rupture disc assembly Thermostat The HTHS Series II is equipped with a thermostat that prevents the instrument from overheating. If temperature rises to an unacceptable level (160°C ± 5°C), power will be removed from the HTHS and will not be restored until the instrument has cooled to a safe state.
  • Page 10: Assembly

    Instrument Company immediately. Damaged items Retain all packing materials until the instrument is connected and func- tioning properly. If any component(s) must be returned to CANNON ® Instrument Company, the damaged item(s) should be packaged in the original shipping container. Refer to the final chapter of this manual for instructions on returning defective equipment.
  • Page 11: Connecting The Syringe Hose Adapter

    500 psi. CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 12: Checking The Power Supply

    3. Flip the PRESSURE/VACUUM switch to VACUUM and check the adapter for a slight suction with your finger.The VPU has been carefully adjusted at CANNON ® but it may be necessary to readjust the force of the vacuum by turning the ADJUST dial on the front panel of the unit.
  • Page 13 This page intentionally left blank. CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 14: High Shear Viscosity Calculator

    The High-Shear Viscosity Calculator is a standalone software package that calculates viscosity of a sample in mPa·s (cP) using efflux time and ® pressure values obtained during testing with the CANNON High- Temperature High-Shear Viscometer (HTHS). The Calculator also provides a routine for ensuring the accuracy of constants used in viscos- ity calculation.
  • Page 15: Program Installation

    RUN window. Click Open. Then click OK. 2. Follow the installation program prompts to complete the installation. The default path for installation is C:\Program Files\CANNON Instrument Company\HTHS. The setup program will install the High-Shear Viscosity Calculator files to the drive and directory specified.
  • Page 16: Hths Pre-Operation Checklist

    HTHS via the rear panel connection (see photo). Listen for signs of gas leakage. HTHS Series II nitrogen connection to internal valve system (panel removed to provide internal detail) Checking solenoid valves 2.
  • Page 17 You may need to practice working with the valve settings to establish the desired pressure in a relatively short time. The internal valve system for the HTHS Series II includes a replaceable CAUTION rupture disc as a safety measure to prevent pressure from rising above 700 psi.
  • Page 18: Standard Hths Test Procedure

    2. Make sure that the waste oil container is in its position at the base of the HTHS under the viscometer cell exit tubes. Cap the HTHS exit tubes to prevent oil leakage during the preliminary part of the test. Satisfactory stoppers are available from CANNON ® (Catalog #9726- N98;...
  • Page 19: Soaking The Test Sample

    DO NOT PROCEED TO THE NEXT STEP OF THE RUN PROCEDURE NOTE IF THE PLUG VALVE FOR THE TEST CELL IS IN THE OPEN (HANDLE VERTICAL) POSITION. The HTHS Series II test cycle will not begin unless the plug valve is closed. 9. Remove the adapter from the charge cylinder.
  • Page 20: Setting Sample Volume

    CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 21: Testing Sample(S)

    CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 22 7. Repeat the test procedure for each cell. CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 23 This page intentionally left blank. CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 24: Daily Maintenance

    To prevent this occurrence, CANNON ® recommends that all five cells be flushed at the conclusion of testing each day with approximately 5 ml of a low viscosity neutral oil (2 mPa·s at 150°C) or commercially-available...
  • Page 25 This page intentionally left blank. CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 26: Calibration

    SAE J300 requires HTHS viscosities from 2.6 to 3.7 mPa•s. Because the HT75 standard has a nominal viscosity of 2.7 mPa•s and the HT150 standard has a nominal viscosity of of 3.7 mPa•s, CANNON ® recommends that the users initially calculate cell volumes using either of these standards.
  • Page 27 This page intentionally left blank. CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 28: Maintenance & Repair

    CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 29 Follow the instructions in APPENDIX A to reassemble and replace the viscometric cell. CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 30: Spare Parts List

    FITTING, MALE CONN B-200-1-4 CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 31 TEST HIGH SHEAR VISCO CALC CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 32: Warranty/Return Information

    WARRANTY/RETURN INFORMATION Products limited warranty ® In addition to other manufacturers’ warranties, CANNON Instrument Company (“the Company”) warrants all products (other than reagents and chemicals) delivered to and retained by their original purchasers to be free from defect in material and workmanship for one year from the date of the Company’s invoice to the purchaser.
  • Page 33: Returning A Product To Cannon

    Shipping notification Products returned without prior notification (by either telephone or fax), or without Cannon’s authorization, will not be accepted. The customer may be billed a testing fee if a product is returned to ® CANNON and found to be working properly.
  • Page 34: Appendix A-Hths Capillary Replacement

    (see photo, next page). CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 35: Removing The Capillary

    CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 36: Reassembling The Cell

    24. Set the selector switch for the correct cell and press the START/ STOP switch. CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 37 HTHS opera- tions. CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 38: High-Temperature High-Shear Rheometer Cell Calibration Record

    Cell Calibration Record CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 39 This page intentionally left blank. CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 40: Appendix B-Hths Rupture Disc Replacement

    The angular seat rupture disc design is rated for up to 700+ psig. NOTE Installation The rupture disc for the HTHS Series II instrument is sold and installed as an assembly that includes the rupture disc itself, vacuum support and holddown ring, conical threaded outlet and inlet connector. The rupture disc assembly is connected inline with the HTHS pressure system via a threaded connection on the inlet hex (see images, next page).
  • Page 41: Preventive Maintenance

    CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 42: Appendix C-Figures & Diagrams

    Plug Valve construction CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
  • Page 43: Viscometric Cell And Exit Capillary (Cutaway View)

    (variable) (variable) (variable) CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...