Testing sample(s) ......................18 CANNON CANNON CANNON CANNON CANNON High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual ® CANNON CANNON CANNON Revision 1.0d—June, 2011; CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
Page 3
SPARE PARTS LIST 10 WARRANTY/RETURN INFORMATION Products limited warranty....................29 Reagent and chemical warranty ..................29 Returning a product to CANNON® ................. 30 APPENDIX A—HTHS CAPILLARY REPLACEMENT Replacing an HTHS capillary ................... 31 Parts/Tools required: ..................... 31 Removing the viscometric cell ................31 Removing the capillary ..................
The HTHS Series II is capable of testing at a temperature of 150°C as specified in ASTM D 5481. For other applications, temperature may be set as low as 25°C above ambient.
(DEC99) requires HTHS viscosities from 2.6 to 3.7 mPa•s. Because the HT75 standard has a nominal viscosity of 2.7 mPa•s and the HT150 ® standard has a nominal viscosity of of 3.7 mPa•s, CANNON recom- mends that the users calculate cell volumes using either of these stan- dards.
Operator safety Please observe the following safety procedures and notices for proper operation of your HTHS Series II instrument. Deviation from the installa- tion, operation or maintenance procedures described in this manual may result in a hazardous situation and may void the manufacturer's warranty.
Page 7
Refer all service and repairs to qualified personnel. • Do not attempt to service the unit beyond the service and/or repair procedures detailed in this manual or other documentation provided by the manufacturer. Contact CANNON ® Instrument Company for all additional service/repair needs.
The container slides into place at the left front of the HTHS Series II. A hinged door provides protection from hot oil. At the rear, an internal fan draws vapors into the exhaust manifold.
HTHS internal rear view including pressure relief rupture disc assembly Thermostat The HTHS Series II is equipped with a thermostat that prevents the instrument from overheating. If temperature rises to an unacceptable level (160°C ± 5°C), power will be removed from the HTHS and will not be restored until the instrument has cooled to a safe state.
Instrument Company immediately. Damaged items Retain all packing materials until the instrument is connected and func- tioning properly. If any component(s) must be returned to CANNON ® Instrument Company, the damaged item(s) should be packaged in the original shipping container. Refer to the final chapter of this manual for instructions on returning defective equipment.
3. Flip the PRESSURE/VACUUM switch to VACUUM and check the adapter for a slight suction with your finger.The VPU has been carefully adjusted at CANNON ® but it may be necessary to readjust the force of the vacuum by turning the ADJUST dial on the front panel of the unit.
Page 13
This page intentionally left blank. CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
The High-Shear Viscosity Calculator is a standalone software package that calculates viscosity of a sample in mPa·s (cP) using efflux time and ® pressure values obtained during testing with the CANNON High- Temperature High-Shear Viscometer (HTHS). The Calculator also provides a routine for ensuring the accuracy of constants used in viscos- ity calculation.
RUN window. Click Open. Then click OK. 2. Follow the installation program prompts to complete the installation. The default path for installation is C:\Program Files\CANNON Instrument Company\HTHS. The setup program will install the High-Shear Viscosity Calculator files to the drive and directory specified.
HTHS via the rear panel connection (see photo). Listen for signs of gas leakage. HTHS Series II nitrogen connection to internal valve system (panel removed to provide internal detail) Checking solenoid valves 2.
Page 17
You may need to practice working with the valve settings to establish the desired pressure in a relatively short time. The internal valve system for the HTHS Series II includes a replaceable CAUTION rupture disc as a safety measure to prevent pressure from rising above 700 psi.
2. Make sure that the waste oil container is in its position at the base of the HTHS under the viscometer cell exit tubes. Cap the HTHS exit tubes to prevent oil leakage during the preliminary part of the test. Satisfactory stoppers are available from CANNON ® (Catalog #9726- N98;...
DO NOT PROCEED TO THE NEXT STEP OF THE RUN PROCEDURE NOTE IF THE PLUG VALVE FOR THE TEST CELL IS IN THE OPEN (HANDLE VERTICAL) POSITION. The HTHS Series II test cycle will not begin unless the plug valve is closed. 9. Remove the adapter from the charge cylinder.
CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
Page 22
7. Repeat the test procedure for each cell. CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
Page 23
This page intentionally left blank. CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
To prevent this occurrence, CANNON ® recommends that all five cells be flushed at the conclusion of testing each day with approximately 5 ml of a low viscosity neutral oil (2 mPa·s at 150°C) or commercially-available...
Page 25
This page intentionally left blank. CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
SAE J300 requires HTHS viscosities from 2.6 to 3.7 mPa•s. Because the HT75 standard has a nominal viscosity of 2.7 mPa•s and the HT150 standard has a nominal viscosity of of 3.7 mPa•s, CANNON ® recommends that the users initially calculate cell volumes using either of these standards.
Page 27
This page intentionally left blank. CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
Page 29
Follow the instructions in APPENDIX A to reassemble and replace the viscometric cell. CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
WARRANTY/RETURN INFORMATION Products limited warranty ® In addition to other manufacturers’ warranties, CANNON Instrument Company (“the Company”) warrants all products (other than reagents and chemicals) delivered to and retained by their original purchasers to be free from defect in material and workmanship for one year from the date of the Company’s invoice to the purchaser.
Shipping notification Products returned without prior notification (by either telephone or fax), or without Cannon’s authorization, will not be accepted. The customer may be billed a testing fee if a product is returned to ® CANNON and found to be working properly.
24. Set the selector switch for the correct cell and press the START/ STOP switch. CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
Page 37
HTHS opera- tions. CANNON CANNON CANNON CANNON CANNON ® High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual Revision 1.0d—June, 2011; CANNON CANNON CANNON CANNON CANNON ® Instrument Company 2139 High Tech Road • State College, PA 16803 • USA...
The angular seat rupture disc design is rated for up to 700+ psig. NOTE Installation The rupture disc for the HTHS Series II instrument is sold and installed as an assembly that includes the rupture disc itself, vacuum support and holddown ring, conical threaded outlet and inlet connector. The rupture disc assembly is connected inline with the HTHS pressure system via a threaded connection on the inlet hex (see images, next page).