Technical Specifications

Name of product
Model
Weight
Dimensions
Power Consumption
Power Source:
AC adapter – UL (Underwriters
Laboratories) Safety Mark
Input
Output
Battery
Battery Life
Output Specifications:
Waveform
Shape
Maximum Output Voltage:
@500Ω
@2kΩ
@10kΩ
Maximum Output Current:
@ 500Ω
@2kΩ
@10kΩ
Pulse Duration:
30
REVITIVE Medic
2469MD
1.5kg
Ø360mm x 75mm (isorocker enabled)
Ø360mm x 100.5mm (isorocker disabled)
5W
100-240V, 50/60Hz, 0.18A.
5.0Vdc, 1.0A
Lithium Ion, 1200mAh, 3.7V
500 charge cycles
Monophasic
Square symmetrical with polarity reversal
Foot: 35V
Body: 19V
Foot: 101V
Body: 54V
Foot: 133V
Body: 68V
Foot: 71mA
Body: 39mA
Foot: 51mA
Body: 27mA
Foot: 13mA
Body: 7mA
Foot: 450 +/- 45µs
Body: 450 +/- 45µs
or 970 +/- 97µs
Frequency:
Net Charge @ 500Ω
Maximum Phase Charge @ 500 Ω
Maximum Current Density
Maximum Power Density @ 500Ω
Burst Mode:
a. Pulses per burst
b. Bursts per second
c. Burst duration
d. Duty Cycle
ON Time
OFF Time
Disposal: At the end of product life do not throw this product into normal household garbage, but
take it to a collection point for the recycling of electronic equipment.
The Remote Control replicates the controls found on the device
Operating Frequencies
Operating Range Distance
Foot: 20 to 43Hz,
Body: 38 to 51Hz
[0.004]mC
Foot: 53µC
Body: 17µC
Foot: 0.032 mA/cm²
Body: 0.100 mA/cm²
Foot: 0.81 mW/cm²
Body: 1.82 mW/cm²
Foot: 195 - 1092
Body: 291 - 921
Foot: 0.11 - 0.29
Body: 0.10 - 0.30
Foot: 1.91 - 8.35s
Body: 1.90 - 6.50s
Foot: 0.56 - 0.89
Body: 0.65 - 0.86
Foot: 1.90 - 8.40s
Body: 1.90 - 6.50s
Foot: 1.00 - 1.50s
Body: 1.00s
38 KH
z
0 to 25m at horizontal
0 to 18m at +/- 30 degree angle
from the horizontal.
31