TM
Service Manual
Gendex GXCB-500
X-ray Tube Assembly
Nominal X-Ray Tube Voltage
Max. Tube Current
X-Ray Tube Nominal Anode Input Power
X-Ray Tube Maximum Anode Heat Content
(HU = Kvp x mA x Time in Seconds)
X-Ray Tube Single Load Rating
Max. X-Ray Tube Assembly Heat Content
Max. Continuous Heat Dissipation of Tube Assy
High Voltage Supply Requirements
Loading Factors Concerning Leakage Radiation:
CT Scan = 120KV 5mA 12 mS x 618 or 309
PAN Scan = 94KV 5mA
15-18
An annual radiation survey may be typically required. This
•
survey is typically required to be submitted to the local
agency or government body.
A phantom or patients may be used for system training.
•
Employees of the facility may not be used for this training.
The system shall be registered with the local agency or
•
government body.
Imaging Sciences International utilizes the SXR 130-15-0.5 X-ray
Tube from Superior X-Ray Tube Company to manufacture our X-
ray head assemblies.
120 kV
7 mA
65 W
30,000 HU
120kV, 5mA
120K HU
65W
120 VAC at 10 amps
< = 25mR/hr @ 1m