How To Set Up And Run The Emc32 Software; Configuring The Mxe - Keysight Technologies X-Series Getting Started Manual

Mxe emi receiver
Hide thumbs Also See for X-Series:
Table of Contents

Advertisement

Using the EMC32 Software

How to Set Up and Run the EMC32 Software

How to Set Up and Run the EMC32 Software
To be compatible with the EMC32 software, the MXE EMI Receiver must
behave like the R&S ESU. The following procedure details the steps necessary
to correctly configure the instrument to emulate the ESU.

Configuring the MXE

Table 3-1
Step
1. MXE connection
2. Set the MXE GPIB
address
3. Set the EMI Receiver
Mode
The EMC32 software does an ID query when it starts up a test, so the MXE should
always be set to the EMI Receiver Mode before setting up the EMC32 software.
4. Set the EMI Receiver
Mode to ID as an ESU
3. Set the MXE to preset
to the EMC Standard
40
Action
— Connect the GPIB cable between
the PC and the MXE
System
/O Config
a. Press
, I
GPIB Address
b. Press
Mode
EMI Receiver
— Press
,
Mode Setup
a. Press
,
Response
R&S ESU
b. Press
For a CISPR test:
Mode Setup
a. Press
,
CISPR
Mode Setup
b. Press
,
Preset To
, and ensure that it is set
CISPR
to
Or
For a MIL test:
Mode Setup
a. Press
,
MIL
Mode Setup
b. Press
,
Preset To
MIL
,
Notes
GPIB
This is what ESU defaults to
,
20
Enter
,
,
The EMI Receiver Mode has the ability to
identify as an ESU, which is required by the
EMC32 software. Most other modes in the
MXE do not have this ability.
Mode IDN
The MXE uses the EMC Standard to choose
which RBW's to make available, and how to
EMC Std
,
couple them.
Note that CISPR is the default at instrument
EMC Std
,
startup, but if you have set MIL to be the
preset and you press the Mode Preset key,
the instrument will preset to MIL.
EMC Std
,
EMC Std
,
X-Series EMI Receiver Getting Started Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

N9038a

Table of Contents