Important information

5.1 Safety information

General recommendations
Read the safety and operating instructions before operating the device.
Retain safety and operating instructions for future reference.
Adhere to all warnings on the device and in the operating instructions manual.
Follow all instructions for operation and use.
Electrical Shock or Fire Hazard
To prevent electric shock or fire hazard, do not remove cover.
No serviceable parts inside. Refer servicing to qualified personnel.
Do not expose this apparatus to rain or moisture.
Modifications to the unit
Do not modify this equipment without authorization of the manufacturer.
Type of protection (electrical)
Monitor with internal power supply: Class I equipment.
Degree of safety (flammable anesthetic mixture)
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or
nitrous oxide.
Non-patient care equipment
Equipment primarily for use in a health care facility that is intended for use where contact with a patient is
unlikely (no applied part).
The equipment shall not be used with life support equipment.
The user should not touch the equipment, nor its signal input ports (SIP)/signal output ports (SOP) and the
patient at the same time.
Power connection – Equipment with internal power supply
Power requirements: The equipment must be powered by the AC mains voltage and must be connected to
a mains socket with earth connection.
The equipment is intended for continuous operation.
Transient over-voltage
If the device is not used for a long time, disconnect it from the AC inlet to avoid damage by transient over-
voltage.
To fully disengage the power to the device, please disconnect the power cord from the AC inlet.
High magnetic environment
The device shall not be used in the high magnetic environment of an MRI scanner.
The installer shall assess the magnetic environment before installation or use of the device.
Connections
Any external connection with other peripherals must follow the requirements of clause 16 of IEC60601-1 3rd.
Ed. or Table BBB.201 of IEC 60601-1-1 for the medical electrical systems.
To maintain compliance with EMC Regulation, use only shielded interface cables for the connection to
peripheral devices.
24
K5902134 /04
Nio Color 2MP 23"