Scheduled Maintenance; Cleaning - Barco Nio Color 2 MP User Manual

23"
Table of Contents

Advertisement

Maintenance

4.1 Scheduled maintenance

About
The Nio Color 2MP 23" does not require any scheduled maintenance or calibration activities. We recommend
to use QAWeb with the Barco default tests and frequencies to calibrate and maintain the display, or to return
the display to a Barco approved maintenance organization. In any case of doubts, please contact Barco
Healthcare.

4.2 Cleaning

WARNING: Unplug the power cable from the mains power input before cleaning the display.
CAUTION: Take care not to damage or scratch the front glass or LCD. Be careful with rings or other
jewelry and do not apply excessive pressure on the front glass or LCD.
CAUTION: Do not apply or spray liquid directly to the display as excess liquid may cause damage
to internal electronics. Instead, apply the liquid to a cleaning cloth.
To clean the display
Clean the display using a sponge, cleaning cloth or soft tissue, lightly moistened with a recognized cleaning
product for medical equipment. Read and follow all label instructions on the cleaning product. In case of doubt
about a certain cleaning product, use plain water.
Possible cleaning solutions:
70% isopropyl alcohol
1.6% aqueous ammonia
Cidex® (2.4% glutaraldehyde solution)
Sodium hypochlorite (bleach) 10%
"Green soap" (USP)
0.5% Chlorehexidine in 70% isopropyl alcohol
Like Cleansafe® optical cleaning liquid
Do not use following products:
Alcohol/solvents at higher concentration > 70%
Strong alkalis lye, strong solvents
Acid
Detergents with fluoride
Detergents with ammonia at higher concentration > 1.6%
Detergents with abrasives
Steel wool
Sponge with abrasives
Steel blades
Cloth with steel thread
22