8.
Tap NEXT.
The DETECTING CARTRIDGE
ü
screen is displayed, followed by the
CARTRIDGE CHANGED screen.
NOTE
Always dispose of used components
such as cartridges, syringes, nee-
dles, infusion sets, and CGM sensors
following the instructions from your
healthcare provider. Wash your hands
thoroughly before and after handling
System components.
9.
Tap START.
The Starting Fill screen
ü
is displayed.
NOTE
The minimum amount of insulin
required to fill tubing after a cartridge
change is 10 units.
10. Tap STOP once three drops of
insulin are seen at the end of
the tubing.
The Stopping Fill screen
ü
is displayed.
11. Verify that drops are seen and
tap DONE.
Customer Technical Support: (877) 801-6901
Make sure the set is disconnected
from your body and that the tubing
is securely connected to the
cartridge.
Press NEXT to fill your tubing
with insulin.
NEXT
Fill Tubing
Fill Tubing
Amount Filled
Fill Tubing
Fill Tubing
Amount Filled
If you see drops at the end of the
tubing, press DONE. If no insulin
drops are noticed, press FILL to
continue filling.
Amount Filled: 15.5 u
FILL
Load
START
0
u
STOP
15.5
u
DONE