Maintenance And Repair Work - Siemens SITRANS LR250 Compact Operating Instructions

Radar transmitters ma/hart
Hide thumbs Also See for SITRANS LR250:
Table of Contents

Advertisement

Service and maintenance

8.3 Maintenance and repair work

Electrostatic charge
Danger of explosion in hazardous areas if electrostatic charges develop, for example, when
cleaning plastic surfaces with a dry cloth.
● Prevent electrostatic charging in hazardous areas.
8.3
Maintenance and repair work
The device is maintenance-free. However, a periodic inspection according to pertinent
directives and regulations must be carried out.
An inspection can include check of:
● Ambient conditions
● Seal integrity of the process connections, cable entries, and cover screws
● Reliability of power supply, lightning protection, and grounds
Maintenance during continued operation in a hazardous area
There is a risk of explosion when carrying out repairs and maintenance on the device in a
hazardous area.
● Isolate the device from power.
- or -
● Ensure that the atmosphere is explosion-free (hot work permit).
Humid environment
Risk of electric shock.
● Avoid working on the device when it is energized.
● If working on an energized device is necessary, ensure that the environment is dry.
● Make sure when carrying out cleaning and maintenance work that no moisture penetrates
50
WARNING
WARNING
WARNING
the inside of the device.
Compact Operating Instructions, 06/2018, A5E33469191-AB