Downloading individual runs from the logger does not delete those runs from
memory or clear space for further runs to be added. The only way to delete
data from the logger's memory is to perform a reset using Insight (p. 41),
and this deletes all the profile-run data stored in the logger.
In some Insight products, individual runs cannot be selected for download:
Downloading data from multiple profile runs stored in the logger (not in some
Insight products; see text). In this example, all profiles except no. 8 are selected for
download; profile no. 2 has already been downloaded, but will now be
downloaded again; the profiles will not be merged.
7. Wait while the data is downloaded to the PC.
Logger stopped due to going over temperature
the data logger's maximum-permitted internal temperature has been exceeded,
and it may have suffered damage. The reason for the excessive temperature –
which may be the result of process operational problems or the use of an
inappropriate thermal barrier – must be resolved before further profile runs
take place; contact Fluke Process Instruments for advice.
A warning message will also be shown if the logger has stopped recording data
In all cases, data recorded up to the point of error will have been preserved.
For recommended actions to take in response to error messages, see p. 61.
8. If enabled, the Select Process dialog or the Select Oven/Furnace/Kiln,
Recipe and Product dialog then appears in order that you may choose a
50
Using Insight Software
instead, all profiles in memory are downloaded.
If you see the message
due to a low battery.