Type T
Measurement
−196°C to 400°C
range
−321°F to 752°F
Accuracy *
±0.5°C below −100°C
(using sample
±0.3°C above −100°C
interval > 0.8 s)
±0.9°F below −148°F
±0.5°F above −148°F
Resolution
0.1°C/0.2°F
Socket color
Brown
(IEC 60584-3)

Logger LEDs

The logger is equipped with two sets of two LEDs:
• Yellow and green/red show the status of the battery.
• Red and green show the status of the logger and its memory.
To see an animated demonstration of all the LED sequences: in Insight,
select Help > LED Sequences; or, in Insight's Help system, select Introduction >
Logger LEDs.

Battery Status LEDs

Yellow
Green/Red
Off
Off
Flashing every
Off
second
Off
RED
Off
GREEN
Off
Flashing RED
once per second
Double-flash
Off
every second
DATAPAQ TP3
*
There will be an additional error of 0.01°C for every
1°C difference between the temperature at which
the logger is operated (i.e. the logger's internal
temperature) and the temperature at which it was
calibrated. For more-detailed accuracy data, contact
Fluke Process Instruments.
Due to continuing product development,
specifications are subject to change
Meaning
Battery has at least 20% of full charge (charger not
connected), or...
Charger is connected but logger is acquiring data (in which
case logger-status green LED will be flashing), or...
Fault with battery or logger.
Battery has 20% or less of full charge (charger not
connected).
Battery charging.
Charging complete (charger connected).
Battery being preconditioned due to being too hot, too cold
or too deeply discharged (see p. 22).
Lithium batteries are being depassivated (see p. 27).
Specifications and Operation