Page 1
PENKO Engineering BV The Leading Experts In Weighing & Dosing SGM700 Digitizer Manual For more information visit www.penko.com - technical support - literature library...
Page 2
If this Instruction Manual is not the processes. The use of non-specified ‘look- correct manual for the Penko product you alike’ substitution parts may result in the are using, call 0031(0)318-525630 for a risk of fire, electrical hazards, or improper replacement copy.
Page 3
Proper relay use and configuration is the responsibility of the user. Do not operate this instrument without the front cover being secured. Refer any installation, operation or servicing issues to qualified personnel. WWW.PENKO.COM Penko is an ETC Company e-mail: info@penko.com...
SGM700 Digitizer 3. Explanation of front keys All keys have different functions depending if you are in weighing or menu mode. Pressing key 1 “short”. In Weighing mode: create a new zero level. In Menu mode: increase value by 1 or move up in menu. Pressing key 1 “long”.
SGM700 Digitizer 4. First use of indicator 4.1 First use of indicator -Configuration menu structure- - - - Fun setpoint function settings (not for SGM710) - - - ACn setpoint action settings (not for SGM710) - - - dAC Analog output settings - - - 485 Local bus communication settings (RS485) - - - Eth...
SGM700 Digitizer 4. First use of indicator 4.2 First use of indicator -Weigher settings- Set up the correct indicator setting (step size and decimal point position). The start Turn the indicator on by connecting it to the power supply. Press key 3 for >2 sec.
Page 16
SGM700 Digitizer First use of indicator -continue- When the maximum net weight value is set succesfully, the following screen is visible: Dis- play step size Press key 1 <2 sec. untill you see Ind 5 and press key 2 <2 sec. Ind 5 to set the display step size.
Page 17
SGM700 Digitizer First use of indicator -continue- Press key 1 to define the point position and confirm by pressing key 3 for >2 sec. Left Right Confirm Done succesfully, the following screen is visible: Press key 3 <2 sec to go back to Configuration Menu.
Page 18
SGM700 Digitizer First use of indicator -continue- 4.3 First use of indicator -Calibration- Press key 3 for >2sec. to get in to Configuration Menu. SGM710) the Calibration parameters by pressing <2 sec. until you see - - - CAL Check and delete calibration points. To enter the Calibration settings, press key 2 <2 sec.
Page 19
SGM700 Digitizer First use of indicator -continue- Calibration Settings -continue- When a number is shown, the deletion of one calibration point is completed and more points need to be deleted. Press key 3 >3 sec to do so. >3 sec. When all calibration points are deleted, the following screen is visible: Entering new calibration points.
SGM700 Digitizer First use of indicator -continue- Calibration Settings -continue- The indicator now shows CP2 to calibrate the gain point (CP2). And will automatically jump to: Use key 1 and key 2 to enter the reference value. Key 1 is used for changing the number (1-9), key 2 is used for changing the position of the cursor.
The SGM can also be configured for first use by using PI mach II software. Download Penko Suite from the PENKO website (http://www.penko.com/db/ downloads/) and install Penko Suite. The following items will be installed: *Pi Mach II Program Interface to configure all Penko devices...
Page 22
If this doesn’t work, make sure other USB devices like mobile phones are disconnected from the PC. In case of Windows 8, a USB driver manual is available in the Penko Suite. In the communication setup it’s also possible to connect the SGM720 through Ethernet.
Page 23
SGM700 Digitizer First use of indicator -continue- First use by using PI Mach II software -continue- Go to Start Quick setup and click the button: The follo- wing tree is shown: Use step 1 to set the Weigher Parameters: The settings you can make are: Unit: set the weigher unit (kg, lbs, T, etc.) this will be shown in PI.
Page 24
SGM700 Digitizer First use of indicator -continue- First use by using PI Mach II software -continue- Use step 2 to select the type of indicator: Set the type of weighing the SGM will be used for. This will automaticly set the most common filter settings to get a stable weigher signal.
Page 25
SGM700 Digitizer First use of indicator -continue- First use by using PI Mach II software -continue- Use step 4 to set the communication (different for every type): SGM720 (Ethernet): Here you can set: Name: here you can give the SGM720a name. Example “platform 1”. Address: set the IP-address for the SGM720.
Page 26
Buslink Subaddress: Set the Buslink Subaddress for the communication between PENKO devices. Baudrate: Set the commication speed of the SGM730 options are: 100k, 125k, 250k, 500k. (the speed of all PENKO devices in the communication bus must be the same). SGM740 (PROFIBUS):...
Page 27
SGM700 Digitizer First use of indicator -continue- First use by using PI Mach II software -continue- SGM750 (Serial): Here you can set (for both RS232 and RS422): Protocol: Set the Protocol that is used on the serial port. Options are: None, Printerm ASCII, NPV Slave, Mudbus RTU, Modbus ASCII.
Page 28
SGM700 Digitizer First use of indicator -continue- First use by using PI Mach II software -continue- Use Done to finish the quick setup. Click End Quick setup: Click Enable Full setup to gain access to all settings (optional): The following tree is shown and gives access to all settings:...
SGM700 Digitizer 5. Main Menu In the Main menu you can change the Setpoints (this option is not available in the SGM710), display the TAC code and display the CAL code. 5.1 Setpoints. (this option is not available in the SGM710) The SGM700 has four outputs that can switch on/off on different levels.
SGM700 Digitizer Main Menu -continue- When done with changing the setpoints press key 3 <2 sec to go back to the main weigher display 5.2 TAC code (Traceable Acces Code). The SGM700 has a TAC code inside. TAC code is number of times the indicator data is changed.
SGM700 Digitizer Main Menu -continue- 5.2 CAL code (Calibration counter). The SGM700 has a CAL code inside. CAL code is the number of times the calibration is changed. When an indicator gets certified this number will be written on the device and is used by the controlling agency to see if the settings aren’t changed after sealing.
SGM700 Digitizer 6. Configuration Menu Press button 3 >2 sec to enter the Configuration Menu. In the Configuration Menu the following options are available: - - - Fun setpoint function settings (not for SGM710) - - - ACn setpoint action settings (not for SGM710) - - - dAC Analog output settings - - - 485...
SGM700 Digitizer Configuration Menu -Fun- 6.1 - - - Fun setpoint function settings (not for SGM710) Configure the weiger mode the outputs has to switch on press key 2 <2 sec to en- ter the setpoint function settings menu. Select the output you want to configure pressing key 1 <2 sec. Fun 1= output 1, Fun 2= output 2, Fun 3= output 3 and Fun 4= output 4.
SGM700 Digitizer Configuration Menu -ACn- 6.2 - - - ACn setpoint action settings (not for SGM710) To set the hysteresis for the outputs press key 2 <2 sec to enter the setpoint function settings menu. Select the output you want to configure pressing key 1 <2 sec. Acn 1= output 1, Acn 2= output 2, Acn 3= output 3 and Acn 4= output 4.
SGM700 Digitizer Configuration Menu -dAC- 6.3 - - - dAC Analog output settings In this menu, all analog output parameters can be set (only for SGM710 or when the analog output option is installed).Options are: dAC 1 Set analog output to minimum level dAC 2 Set analog output to maximum level dAC 3...
Page 37
SGM700 Digitizer Configuration Menu -dAC- -continue- In dAC 2 you can set the analog output to its maximum level for testing purpo- se.Press key 2 <2 sec to set the analog output to maximum. Press key 3 <2 sec to go back. following screen is vissable: In dAC 3 you can set the analog output to a level...
Page 38
SGM700 Digitizer Configuration Menu -dAC- -continue- The options are: Weigher 10 Weigher x 10 Fast gross 11 Fast gross x 10 Fast net 12 Fast Net x 10 Display Gross 13 Display Gross x 10 Display Net 14 Display Net x 10 Tare 15 Tare x 10 Peak...
Page 39
SGM700 Digitizer Configuration Menu -dAC- -continue- The following screen will be vissable: In dAC 6 you set the weigher end value for the analog output. At this value the analog output stops with its maximum value. Press key 2 < 2sec. to enter dAC 6. Fill in the wanted end value using key 1 and key 2.
SGM700 Digitizer Configuration Menu -dAC- -continue- Options are: 1 RAW Register value 65535 parts 2 0-24mA input value 0 to 24 mA will be calculated from 0 to 100,00% 3 0-20mA input value 0 to 20 mA will be calculated from 0 to 100,00% 4 4-20mA input value 4 to 20 mA will be calculated from 0 to 100,00% 5 4-24mA...
Ethernet settings (SGM720 only) Protocols that can be used are Ethernet IP, Omron Fins and Modbus TCP. For protocol description please download PENKO Suite from www.penko.com In this menu, the communication settings can be set for the ethernet port. Options are:...
Page 42
SGM700 Digitizer Configuration Menu -Eth- -continue- In Adr 1 you set the first three numbers of the IP address you want to give the SGM720 (example: 192.168.151.112). Use key 1 and key 2 to change the number Key 1 is for changing the number (1-9), key 2 is for changing the position of the cursor.
Page 43
SGM700 Digitizer Configuration Menu -Eth- -continue- When confirmed the following screen will be visible: In Adr 4 you set the fourth three numbers of the IP address you want to give the SGM720 (example: 192.168.151.112). Use key 1 and key 2 to change the num- ber Key 1 is for changing the number (1-9), key 2 is for changing the position of the cursor.
Page 44
SGM700 Digitizer Configuration Menu -Eth- -continue- Increase Decrease Left Right Confirm When confirmed the following screen will be visible: In Sub 3 you set the third three numbers of the Subnet address you want to give the SGM720 (example: 255.255.255.000). Use key 1 and key 2 to change the number Key 1 is for changing the number (1-9), key 2 is for changing the position of the cursor.
Page 45
SGM700 Digitizer Configuration Menu -Eth- -continue- In gAt 1 you set the first first numbers of the Gateway address you want to give the SGM720 (example: 192.168.001.001). Use key 1 and key 2 to change the number Key 1 is for changing the number (1-9), key 2 is for changing the position of the cursor.
Page 46
Confirm by pressing key 3 >2 sec. Increase Decrease Left Right Confirm When confirmed the following screen will be visible: Starting at firmware version 1.5.0.9.0.1 the PENKO ASCII protocol is available over TCP at port 23.
Configuration Menu –Can- 6.6 - - - Can Canbus settings (SGM730 only) The protocol that can be used is PENKO CAN Buslink (CANopen is not supported). For protocol description please download PENKO Suite from www.penko.com In this menu, the communication settings can be set for profibus.
Page 48
SGM700 Digitizer Configuration Menu –Can- -continue- In CAn 2 you set the Canbus address of the SGM730. Up to 8 devices can com- municate with eachother, sharing inputs, outputs, markers and indicators. Press key 2 <2 sec. to enter CAn 2. Set the address using key 1 and key 2 confirm by pressing key 3 >2 sec.
Page 49
SGM700 Digitizer Configuration Menu –Can- -continue- In CAn 4 you set the Canbus Baudrate of the SGM730. Set the communication speed using key 1 and key 2 confirm by pressing key 3 >2 sec. (options are: 1= 100kbps, 2=125kbps, 3=250kbps and 4=500kbps). Increase Decrease Confirm When confirm the following screen will be vissable:...
SGM700 Digitizer Configuration Menu –Pb- 6.7 - - - - Pb Profibus settings (SGM740 only) For GSD file and protocol description please download PENKO Suite from www.penko.com. In this menu, the communication settings can be set for profibus. Options are:...
Page 51
SGM700 Digitizer Configuration Menu -Pb- -continue- In Pb 2 you set the profibus Value mode. The profibus value can be shown as Integer (direct value without decimal point) or as Floating Point (real value with decimal point). Press key 2 <2 sec. to change the mode. oprions are: Pb2 FL Floating point Pb2 In Integer...
232 1. Set the Prot using key 1 and key 2 confirm by pressing key 3 >2 sec. (options are: 1= None, 2= Printer, 3= ASCII, 4= NPV Slave, 5= Modbus-RTU, 6= Modbus ASCII). For protocol descriptions please download PENKO Suite from www.penko.com...
Page 53
SGM700 Digitizer Configuration Menu –232- -continue- When confirmd the following screen will be vissable: In 232 2 you set the RS232 address of the SGM750. Press key 2 <2 sec. to enter 232 2 Set the address using key 1 and key 2 confirm by pressing key 3 >2 sec. (options are: 0-255).
Page 54
SGM700 Digitizer Configuration Menu –232- -continue- In 232 4 you set the RS232 Parity of the SGM750. Press key 2 <2 sec. to enter 232 4 Set the Parity using key 1 and key 2 confirm by pressing key 3 >2 sec. (options are: 1= none, 2= odd, 3= even, 4= mark, 5=space).
Page 55
SGM700 Digitizer Configuration Menu –232- -continue- In 232 6 you set the RS232 ASCII Indicator of the SGM750. This setting only ta- kes effect when using the ASCII protocol. Press key 2 <2 sec. to enter 232 6 Set the Indicator numer that you want to sent out over the RS232 port by using key 1 and key 2 confirm by pressing key 3 >2 sec.
422 1. Set the Prot using key 1 and key 2 confirm by pressing key 3 >2 sec. (options are: 1= None, 2= Printer, 3= ASCII, 4= NPV Slave, 5= Modbus-RTU, 6= Modbus ASCII).For protocol descriptions please download PENKO Suite from www.penko.com...
Page 57
SGM700 Digitizer Configuration Menu –422- -continue- When confirmd the following screen will be vissable: In 422 2 you set the RS422 address of the SGM750. Press key 2 <2 sec. to enter 422 2 Set the address using key 1 and key 2 confirm by pressing key 3 >2 sec. (options are: 0-255).
Page 58
SGM700 Digitizer Configuration Menu –422- -continue- In 422 4 you set the RS422 Parity of the SGM750. Press key 2 <2 sec. to enter 422 4 Set the Parity using key 1 and key 2 confirm by pressing key 3 >2 sec. (options are: 1= none, 2= odd, 3= even, 4= mark, 5=space).
Page 59
SGM700 Digitizer Configuration Menu –422- -continue- In 422 6 you set the RS422 ASCII Indicator of the SGM750. This setting only ta- kes effect when using the ASCII protocol. Press key 2 <2 sec. to enter 422 6 Set the Indicator numer that you want to sent out over the RS424 port by using key 1 and key 2 confirm by pressing key 3 >2 sec.
SGM700 Digitizer Configuration Menu -Ind- 6.10 - - - Ind Indicator settings In this menu, the Indicator settings can be set. Options are: Ind 1 Maximum display value Ind 2 No motion band Ind 3 Stable time Ind 4 Digital overall filter Ind 5 Display step size Ind 6...
Page 61
SGM700 Digitizer Configuration Menu -Ind- -continue- The following screen is vissable: Use key 1 and 2 to change the value. Key 1 is for changing the number (1-9), key 2 is for changing the position of the cursor. Confirm by pressing key 3 for >2 sec.
Page 62
SGM700 Digitizer Configuration Menu -Ind- -continue- The following screen is vissable: In Ind 3 you set the Stable time Indicator gives stable signal when weigher value is stable within the range set in Ind 2 and time set with Ind 3. Choose a value between: 000.000 –...
Page 63
SGM700 Digitizer Configuration Menu -Ind- -continue- To change the filter press key 2 < 2 sec. The following screen is vissable: Use key 1 change the filter. Confirm by pressing key 3 for >2 sec. Down Confirm The following screen is vissable: In Ind 5 you set the display step size.
Page 64
SGM700 Digitizer Configuration Menu -Ind- -continue- Use key 1 to select the correct step size. Choose between 1, 2, 5, 10, 20, 50 and confirm by pressing key 3 for >2 sec. Example step size: weigher value is 2005 kg Down Confirm Step Size Weight (kg)
Page 65
SGM700 Digitizer Configuration Menu -Ind- -continue- The following screen is vissable: In Ind 7 you set the Display refreshment speed. The Display refreshment speed defines the times the weigher value is refresht per second. Options are: 1, 2, 3, 5, 10, 25, 50.
Page 66
SGM700 Digitizer Configuration Menu -Ind- -continue- In Ind 8 you set the operation mode of the SGM700. Set the operation mode of the unit to Industrial or Certified. In Industrial mode it’s always possible to change the indicator parameters and calibration. In certified mode the unit will be sealed by marks and also the weighing parameters will be blocked to satisfy to the cali- bration laws.
SGM700 Digitizer Configuration Menu -Ind- -continue- To change the sample rate press key 2 < 2 sec. The following screen is vissable: Press key 1 to choose the sample rate and confirm by pressing key 3 for >2 sec. Down Confirm The following screen is vissable: Configura- tion Menu -rng-...
Page 68
SGM700 Digitizer Configuration Menu -rng- -continue- The following screen is vissable: In rng 1 you set the number of display divisions. Set the number of divisions when the indicator has to display with the next step size. Auto ranging starts with step size set at Ind 5 and is disabled when range size is set to 0.
Page 69
SGM700 Digitizer Configuration Menu -rng- -continue- To change the maximum step size press key 2 < 2 sec. The following screen is vissable: Use key 1 change the filter. Key 1 is for changing the number. Confirm by pressing key 3 for >2 sec. Down Confirm Example Max Step: If the settings are:...
SGM700 Digitizer Configuration Menu -rng- -continue- In rng 3 you set the auto range reset option. Choose between: oF: Multi Range = the highest shown step size will be reseted after the signal has been lower or equal to zero on: Mulit Interval = the highest shown step size will be reseted after the signal reaches the previous range.
Page 71
SGM700 Digitizer Configuration Menu -FIL- -continue- The following screen is visible: In FIL 1 you can set the display filter band. Set the band where the filter is active. This parameter works together wit FIL 2. Choose a value between –99999 kg and 999999 kg.
Page 72
SGM700 Digitizer Configuration Menu -FIL- -continue- The following screen is visible: Use key 1 change the filter. Confirm by pressing key 3 for >2 sec. Down Confirm The following screen is visible: In FIL 3 you set the band within the indicator will show 0.
Page 73
SGM700 Digitizer Configuration Menu -FIL- -continue- In FIL 4 you can set the number of the indicator shown in the display. Choose a number between 1 and 19. When using the FIL parameters use 4, 5, 12 or 13. The following screen is visible: The options are: Weigher...
SGM700 Digitizer Configuration Menu -dSF- 6.13 - - - dSF Digital filter settings In this menu, the digital filter settings can be set. This filter is a 2 order filter. The filter effects all signals up to and including the cutoff frequency. Options are: dSF 1 Filter type dSF 2...
Page 75
SGM700 Digitizer Configuration Menu -dSF- -continue- The following screen is visible: In dSF 2 you can set the Cutoff frequency. Determines the range used for filtering the signal. Choose between oFF, 1.0Hz, 1.4Hz, 2.5Hz, 5.0Hz, 10.0Hz, 20.0Hz and 40Hz. Press key 2 <2 sec. to change the cutoff frequency range. The following screen is visible: Use key 1...
SGM700 Digitizer Configuration Menu -PCL- Use key 1 and 2 to change the value. Key 1 is for changing the number (1-9), key 2 is for changing the position of the cursor. Confirm by pressing key 3 for >2 sec. Down Left Right...
Page 77
SGM700 Digitizer Configuration Menu -PCL- -continue- In PcL 1 you can set the polaritiy of the input range. Un=Unipolar mode the input range for load cells is –0.2 mV/V to + value set at Range. Bi=Bipolar mode the input range for load cells is –value set at Range to + value set at Range.Press key 2 <2 sec.
Page 78
SGM700 Digitizer Configuration Menu -PCL- -continue- Use key 1 change the sensitiviy. Confirm by pressing key 3 for >2 sec. Down Confirm The following screen is visible: In PcL 3 you can set the Input offset. Choose bet- ween choose a sample value between –50000 and 50000. Press key 2 <2 sec. to change the in- put offset The following screen is visible:...
SGM700 Digitizer Configuration Menu -PCL- -continue- The following screen is visible: The following screen is visible: To recall the factory settings press key 3 >2 sec. cancel press key 3 <2 sec. Cancel Recall The following screen is visible: Configuration Menu -CAL- 6.15 - - - CAL Calibration settings In this menu, the Calibration settings can be set.
Page 80
SGM700 Digitizer Configuration Menu -CAL- -continue- Press key 2 <2 sec to enter the calibration settings. The following screen is visible: In CAL 1 you can set the calibration points for the weigher. Press key 2 <2 sec. to set the calibration points. After entering, the following screen is visible (if there are no calibration points available):...
Page 81
SGM700 Digitizer Configuration Menu -CAL- -continue- The indicator now shows CP2 to calibrate the gain point (CP2). And will automatically jump to: Use key 1 and key 2 to enter the reference value. Key 1 is used for changing the number (1-9), key 2 is used for changing the position of the cursor.
Page 82
SGM700 Digitizer Configuration Menu -CAL- -continue- Use key 1 <2 sec. to toggle between the actual weight and the actual weight x10. Use key 2 <2 sec. to toggle between the actual weight and the ADC value. When finished press key 3< 2 sec. following screen is visible: Use CAL 3 to check and delete all existing cali-...
Page 83
SGM700 Digitizer Configuration Menu -CAL- -continue- When a number is shown, the deletion of one calibration point is completed and more points need to be deleted. Press key 3 >3 sec to do so. >3 sec. When all calibration points are deleted, the following screen is visible: In CAL 4 you can set a Deadload compensation.
SGM700 Digitizer Configuration Menu -CAL- -continue- Normally, the deadload is zero, but it is possible to change the line position if there is weight on the scale. To do so, edit the actual weigh value to the new known value. When the new dead load is set the follo- wing screen is visible:...
Page 85
SGM700 Digitizer Configuration Menu -tCL- -continue- Press key 2 <2 sec to enter the theoreatic calibration settings. The following screen is visible: In tcL 1 you can set the maximum loadcell load. When more then one loadcell is used all loadcells should have the same maximum load. Press key 2 <2 sec. to set the maximum load.
Page 86
SGM700 Digitizer Configuration Menu -tCL- -continue- The following screen is visible: Use key 1 and key 2 to enter the sensitivitiy of loadcell 1. Key 1 is used for changing the number (1-9), key 2 is used for changing the position of the cursor.
Page 87
SGM700 Digitizer Configuration Menu -tCL- -continue- Use key 1 and key 2 to enter the sensitivitiy of loadcell 2. Key 1 is used for changing the number (1-9), key 2 is used for changing the position of the cursor. Sensitivity has to be filled in as 0.00000 mV/V. Press key 3 >2 sec. to set the sensitivity of the loadcell.
Page 88
SGM700 Digitizer Configuration Menu -tCL- -continue- The following screen is visible: In tcL 5 you can set the loadcell sensitivity for loadcell 4. This information can be found on the datasheet deliverd with the loadcell. Press key 2 <2 sec. to set the loadcell sensitivity.
SGM700 Digitizer Configuration Menu -gCL- 6.17 - - - gCL Geographic calibration In this menu, the Geographic calibration settings can be set. Here you can set a geographic information of the loadcells filled in al tCL. Options are: gCL 1 Origin latitude gCL 2 Origin elevation...
Page 90
SGM700 Digitizer Configuration Menu -gCL- -continue- Down Left Right Confirm The following screen is visible: In gcL 2 you can set the origin elevation of the loadcell. This is the geographic elevation of where the loadcell is manufactured. Press key 2 <2 sec. to set the origin elevation. The following screen is visible: Use key 1 and...
Page 91
SGM700 Digitizer Configuration Menu -gCL- -continue- In gcL 3 you can set the location latiude of the loadcell(s). This is the geographic latitude of where the loadcell is going to be used. Press key 2 <2 sec. to set the location latitude. The following screen is visible: Use key 1 and key 2 to enter the location latitude of...
SGM700 Digitizer Configuration Menu -gCL- -continue- Use key 1 and key 2 to enter the location elevation of the loadcell(s). Key 1 is used for changing the number (1-9), key 2 is used for changing the position of the cursor. And press key 3 >2 sec.to set the location latitude. Choose value between –1000 and 30000 meter Down Left...
SGM700 Digitizer Configuration Menu -CLo- -continue- The following screen is visible: Set the time. Use key 1 and key 2 to change the number Key 1 is for changing the number (1-9), key 2 is for changing the position of the cursor. Confirm by pressing key 3 >2 sec.
SGM700 Digitizer Configuration Menu -rcl- -continue- To set all parameters back to factory settings Press key 3 >2 sec. The device will reboot. To reset the file system Press key 1 >2 sec. The following screen is visible. Confirm by pressing key 3 >2 sec. The device will reboot.
SGM700 Digitizer 7. Firmware update Connect the SGM to the computer through USB. Start PI Mach II. Set communi- cation to USB. Start the Firmware Update Manager. Click Open and select the PIP file. Click Search For Devices and select the device with source “0”. double click or the arrow button to put the address in the Destination List and click OK.
Page 96
SGM700 Digitizer Firmware update -continue- Now set the SGM in boot mode: Go into the configuration menu by pressing key 3 >2 sec. The following screen will appear: Go to- - - SoF by pressing key 1 <2 sec untll you see - - - SoF Press key 2 <2 sec to enter Boot mode.
SGM700 Digitizer 10. Profibus Protocol Description Inputs to PLC D word 32 bit Weight register word 16 bit Status byte 8 bit Reserved byte 8 bit Weight selected register Outputs from PLC word 16 inputs Input 1-16 byte 8 bit Command word 16 outputs Output 201-216...
Page 100
SGM700 Digitizer 11. Standard Factory Settings Description Display Value Your setting Setpoint function Fun 1 Fun 2 Fun 3 Fun 4 Setpoint action Acn 1 000,010 Acn 2 000,010 Acn 3 000,010 Acn 4 000,010 Analog output dAC 4 dAC 5 000.000 dAC 6 010.000...
Page 101
SGM700 Digitizer 11. Standard Factory Settings -continue- Description Display Value Your setting Ethernet gAT 1 gAT 2 gAT 3 gAT 4 Indicator Ind 1 10.009 Ind 2 Ind 3 1.000 Ind 4 Ind 5 Ind 6 - - - . - - - Ind 7 Ind 8 Ind 9...
SGM700 Digitizer Appendix I Description Definition Weight filtered net weigher value that can react on mulit range/interval Fast Gross unfiltered gross weigher value Fast Net unfiltered net weigher value Display Gross filtered gross weigher value Display Net filtered net weigher value Tare tare value Peak...