Frequently Asked Questions (FAQs)

Frequently Asked Questions

Frequently Asked Questions
1 What can I do if I observe a leap in the HDR signal?
2 How are the signals named?
3 Can I use additional detectors if I include a UIB II to the HDR-DAD
4 Which firmware is requiered to run the HDR-DAD solution?
48
Wrong delay calibrations cause leaps in the data signal (see
page 48).
Double check the delay calibration between the 2 detectors.
Figure 26
Leaps in the HDR signal
DAD1 = combined HDR-DAD signal
LNG1 = signal of the single DAD with the long Max-Light Cartridge Cell
installed
SHT1 = signal of the other single DAD with the short Max-Light Cartridge
Cell installed
solution?
It is not recommended to use additional detectors with the HDR-DAD
solution.
You need firmware revision B.06.57 or later to run the HDR-DAD solution.
Figure 26
on
HDR-DAD Solution