Setting Up The Measuring Points; Establishing The Measuring Range - Siemens SITRANS LH100 Compact Operating Instructions

Pressure transmitter; transmitter for hydrostatic level
Hide thumbs Also See for SITRANS LH100:
Table of Contents

Advertisement

Installing/mounting

4.4 Establishing the measuring range

4.3

Setting up the measuring points

Figure 4-2
4.4
Establishing the measuring range
Calculating the measuring range with process media with a density ≠ 1000 kg/m3 (process medium ≠
water)
p = ρ * g * h
with:
ρ = density of the process medium
g = local gravitational acceleration
h = maximum level
16
Cable box
Anchor clamp
Basic procedure for setting up the measuring points
Pressure transmitter
SITRANS LH100 Transmitter for Hydrostatic Level
Compact Operating Instructions, 01/2019, A5E32588610-AA

Hide quick links: