BMA-2 Operating Manual Your humimeter BMA-2 at a glance The main unit Name Control panel and display Measuring chamber Drawer USB port Power supply Page 2...
The control panel Name Print button Save button Control buttons for display Start button On/off button Display LED light solid blue: ready for use flashing blue: measurement in progress flashing red: fault (see"11. Faults") Page 3...
BMA-2 Operating Manual The display Name Temperature Moisture content in % (see "6.1 How moisture defined") Display symbols Atro weight/m³ (optional) Bulk density (optional) Product type The display symbols Symbol Name Symbol Name Enter Change input level Down Back...
The menus The device has two different menus: product selection menu and main menu: Product selection menu Name Open main menu Show the last recorded values For changing the product type Main menu The main menu comprises the following menu items: •...
BMA-2 Operating Manual Table of contents Your humimeter BMA-2 at a glance �������������������������������������������� 2 The main unit ....................2 The control panel ....................3 The display ......................4 The display symbols ..................4 The menus ......................5 1� Introduction �������������������������������������������������������������������� 9 Information about this operating manual ..........
Page 7
Taking a measurement ..............16 Switching the device off ..............16 5� The measuring process ������������������������������������������������� 17 Preparing a measurement ............... 17 Taking a measurement ..............17 Saving individual readings ............... 19 Saving several readings (a measurement series) at the same time ..20 Printing measured values ..............
Page 8
BMA-2 Operating Manual Deactivating options ............... 40 Selecting °C/°F ................40 Reducing the device's power consumption ........41 Configuring the material calibration function ........42 Changing the password ..............42 9.10 Resetting the device to its factory settings ........43 10�...
Introduction Information about this operating manual This operating manual is designed to enable you to use the humimeter BMA-2 safely and effectively. It is part of the device, has to be stored nearby and must be easily ac- cessible to users at all times.
BMA-2 Operating Manual Symbols used in this manual All of the safety information provided in this manual is shown with a corresponding symbol. WARNING It is essential to observe this warning. Non-compliance can lead to serious irreversible or fatal injury.
For your safety The device complies with the following European directives: • Restriction of Hazardous Substances in Electrical and Electronic Equipment (RoHS) • Electrical equipment (Low Voltage Directive, LVD) • Electromagnetic Compatibility Directive (EMC) • Machinery The device corresponds to state-of-the-art technology. However, it is still associated with a number of residual hazards.
BMA-2 Operating Manual Danger from electric current WARNING Electric shock Danger to life due to electric shock when operating with a defective mains cable or defective device. ► Disconnect the device from the power supply and secure it against being switched on again.
Data package, ATRO package) • Determination of bulk density and ATRO dry weight (tons/m³) for humimeter BMA-2 (Included in the ATRO equipment package) • External rechargeable battery - Described in a separate operating manual • Bluetooth module - Described in a separate operating manual •...
BMA-2 Operating Manual Setting up the device Note the following: • Do not leave the device out in the rain. The device and measuring chamber are not waterproof. • Do not position the device in damp rooms.. • Place the device on a stable, level and dry surface.
Using the device - Basics Inserting the measuring chamber Open the drawer of the device by lift- ing and pulling the handle towards you (figure 1). Hang the empty measuring chamber in the drawer. » The handle and the suspension of the measuring chamber must hang in the plastic brackets of the drawer (figure 2).
BMA-2 Operating Manual Automatic drawer check » After switching on, the device goes through a sen- sor check of the drawer. The display will show the message Check Door (figure 5). • Open and close the drawer. Automatic calibration »...
The measuring process Preparing a measurement To do so: The display has to show the measuring window (figure 9). Open the drawer of the device by lifting and pulling the handle towards you (figure 10). Take out the measuring chamber. Completely fill the measuring chamber with the material being measured.
Page 18
BMA-2 Operating Manual Keep pressed to scroll through the types and confirm by pressing Press to start the measurement. » The LED is blinking in blue during the entire mea- surement. » The display shows the active measurement process (figure 14).
Saving individual readings To do so: A measurement has been taken. The display will now appear as shown in figure 16. Press » The display will now appear as shown in figure 17. Press to enter a name for the saved reading and to finish the measuring process.
BMA-2 Operating Manual Saving several readings (a measurement series) at the same time Take several readings of the same material (see "5. The measuring process"). Press after each measurement. » The display will now appear as shown in figure 20.
» The display will show the following information: Name Name of the measurement series (editable) Bulk density [kg/m³] (average) (optional) Dry weight (atro)/m³ [kg/m³] (average) (optional) Temperature (average) Moisture content (average) Printing measured values To do so: The integrated printer and a paper roll is required (figure 23).
Page 22
BMA-2 Operating Manual 5.5.2 Printing already saved readings To do so: You must have taken and saved one or several readings. Press Select Print Logs (figure 27). To do so, press and confirm by pressing Select Last Log (figure 28) to print the last recorded measurement series.
Viewing individual readings To do so: You must have saved a reading (e.g. 1 log). The display will now show Press Select the required reading. To do so, press » The display will now appear as shown in figure 31. Press to leave this screen.
BMA-2 Operating Manual Deleting all measured values (data log) To do so: You must have taken and saved one or several readings. Press Select Edit Logs (figure 35). To do so, press and confirm by pressing Select Clear Logs (figure 36). To do so, press and confirm by pressing »...
BMA-2 Operating Manual Definition of wood chips types (Norm EN ISO 17225-1) The given numbers, in accordance with EN ISO 17225-1, refer to the particle sizes that fit through the round screen openings. • P16 at least 75% of the mass between 3.15 and 16mm •...
Page 27
6.3.3 Softwood chips For wood chips with fine fraction, mainly (more than two thirds) consisting of soft- wood (spruce, fir, pine, larch). The fine fraction mainly derives from barks, small branches and bushes. For softwood chips sizes from P16 to P45. See example pictures and 45.
For determining the bulk density according to norm EN 14961 a round bucket is used. As the humimeter BMA-2 device uses a rectangular measuring chamber, a compensa- tion factor is stored in the device. This compensation factor has been optimised for wood chips and therefore can differ when measuring other material types.
Exporting measuring values to a computer To do so: LogMemorizer must be installed. And you must have taken and saved one or several moisture readings. Options: You can export moisture readings from the humimeter BMA-2 and initiate the export at your computer. Page 34...
Page 35
BMA-2 will now be sent to your computer. Initiating the data export at your computer Connect the humimeter BMA-2 to your computer using the supplied USB cable: Insert the USB Mini B connector into the humi- meter BMA-2 (figure 55).
Page 36
Import most recent manual log (for importing the most recent manually saved logs). Name Import all manual logs Import most recent manual » The measuring values saved on the humimeter BMA-2 will now be sent to your computer. Page 36...
Checking the device's status Press Select Status. To do so, press and confirm by pressing » The display will then show the status indicator humimeter. » The display will show the following information: Name Serial number Software version Battery status Memory status Confirm by presssing Press...
BMA-2 Operating Manual Configuring the device Turning on Bluetooth The information on Bluetooth is provided in a separate operating manual. Adjust the date/time Press Select Options. To do so, press and confirm by pressing Select Date/Time. To do so, press and confirm by pressing »...
Selecting a language Press Select Options. To do so, press and confirm by pressing Select Language. To do so, press and confirm by pressing Navigate to the required language. To do so, press and confirm by pressing » The settings have been saved. Press to leave the Options menu.
BMA-2 Operating Manual » The settings have been saved. » The °C/°F, BL On Time, Auto Off Time, Materialcalib., Password, Reset options are now activated. Press to leave the Options menu. Press to leave the main menu. Deactivating options Once the device has been switched restarted, the °C/°F, BL On Time, Auto Off Time,...
Reducing the device's power consumption 9.7.1 Configuring the display illumination time To do so: All of the options must be activated (see "9.4 Activating options"). Press Select Options. To do so, press and confirm by pressing Select BL On Time. To do so, press and confirm by pressing Select the required display illumination period (30 seconds/2 minutes/5 minu- tes/10 minutes).
BMA-2 Operating Manual Configuring the material calibration function The type calibration function is described in a separate operating manual. Changing the password To do so: All of the options must be activated (see "9.4 Activating options"). Press Select Options. To do so, press and confirm by pressing Select Password.
9.10 Resetting the device to its factory settings To do so: All of the options must be activated (see "9.4 Activating options"). Press Select Options. To do so, press and confirm by pressing Select Reset. To do so, press and confirm by pressing »...
BMA-2 Operating Manual 10.2 Cleaning the device ATTENTION Do not clean with fluids Water or cleaning fluid getting inside the device can destroy the device. ► Only clean with dry materials. ATTENTION Damage due to improper cleaning Damage to the golden contacts on the weighing plate (figure 64) can destroy the device.
10.3 Replacing the printer paper roll To do so: The integrated printer and a paper roll is required. Both are included in the equipment pack- ages: Data package or ATRO package. Lift the tab of the printer all the way up (figure 66).
BMA-2 Operating Manual 11. Faults If the measures listed below fail to remedy any faults or if the device has faults not listed here, please contact Schaller GmbH. Fault Cause Remedy Measuring error The temperature of the ma- The temperature of the...
Page 47
Fault Cause Remedy Measuring chamber is not Put the filled measuring in the device while taking a chamber into the device. Press to confirm the measurement error message. Measuring chamber empty Put the filled measuring while taking a measurement chamber into the device. Press to confirm the error message.
BMA-2 Operating Manual 12. Troubleshooting 12.1 Cleaning the contacts of the measuring chamber To do so: The automatic calibration is not successful and the exclamation mark on the display does not go away. The contacts of the measuring chamber are polluted.
13. Transportation, storage and disposal 13.1 Transporting the device ATTENTION Damage due to improper transport The device can be damaged or ruined by a dispatch with various parcel services or by post. ► Only ship the device in its original packaging. ►...
BMA-2 Operating Manual 14. Device information 14.1 EC declaration of conformity DECLARATION OF CONFORMITY______ in accordance with the following Directives: Low Voltage Directive 2014/35/EU, EC Machinery Directive 2006/42/EC, Electromagnetic Compatibility (EMC) Directive 2014/30/EC, Schaller GmbH Max-Schaller-Straße 99 A – 8181 St. Ruprecht hereby declare that the following product type humimeter BMA ;...
14.2 Equipment packages In addition to the standard version without any accessories, the device is available in three further, different equipment variants: USB package • USB flash drive with LogMemorizer measuring data recording and analysing software and USB cable Data package •...
Page 52
Paper / Board Schaller Messtechnik develops, produces and sells professional moisture meters and turnkey solutions. Schaller GmbH Max-Schaller-Straße 99, A - 8181 St. Ruprecht an der Raab Tel +43 (0)3178 - 28899 , Fax +43 (0)3178 - 28899 - 901 info@humimeter.com, www.humimeter.com...