Assembly
1.
Make sure that the castors have their brakes on.
2.
Connect the base frame to the mains.
3.
Run the bed to the marked height level so that the cams (see picture 3, page 15) on the base frame's
guide way and the shear arms are opposite each other.
4.
Lift the foot end of the top frame so that the two 2 lock
fittings/snap locks fit onto the pins on the shearing arms. (4.1)
Make sure that the locking pin has clicked into place and is
locked (4.2). Then move the foot end of the top frame into
horizontal position.
5.
Place the head end of the top frame in front of the base frame and
lower it down so that the plastic gliders slide into the guide way.
6.
Lift the head end of the top frame and push it into the inserts
until it stops.
7.
Screw the 2x2 finger screws into the side bars.
8.
Connect the plugs of the backrest, thigh and leg motors to
the control unit, see section 7 regarding wiring.
9.
Mount the accessories.