Technical Specifications - ResMed H5i User Manual

Heated humidifier
Hide thumbs Also See for H5i:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
368148 H5i User Guide EUR1 EPN Multi.fm Page 4 Thursday, November 26, 2009 8:52 AM
Problem/Possible cause
Water splashing on your face from the H5i
The water tub is overfilled.
Condensation is forming in the air tube and mask.
Leaking water tub
The water tub may be damaged or cracked.
The cleanable water tub is not assembled
correctly.
Air feels too warm/cold in the mask
The temperature of the ClimateLine air tubing is
set too high/low.

Technical specifications

Maximum heater plate temperature
Temperature cut-out
Maximum gas temperature
Nominal dimensions (L x W x H)
Weight (Water tub)
Weight (Cleanable water tub)
Water capacity
Docking station
Cleanable water tub
Water tub
Operating temperature
Operating humidity
Storage and transport temperature
Storage and transport humidity
ClimateLine air tubing
Heated tubing temperature cut-out
IEC 60601-1 classification
Electromagnetic compatibility
Notes:
The manufacturer reserves the right to change these specifications without notice.
The temperature and relative humidity settings displayed for ClimateLine are not measured values.
Check with your clinician/service provider before using the SlimLine air tubing with devices other than the S9 or H5i.
4
Solution
Check that the water level is below the maximum water level mark.
Turn the H5i or relative humidity setting down via the menu.
Contact your service provider for a replacement.
Check for damage and reassemble the cleanable water tub correctly.
Turn up/down the heated tubing temperature via the menu.
°
65
C
°
74
C
°
≤41
C
Docking station and water tub: 153 mm x 145 mm x 86 mm
Docking station and unfilled water tub 0.67 kg
Docking station and unfilled water tub 0.77 kg
To maximum fill line 380 mL
Flame retardant engineering thermoplastic, aluminium
Injection molded plastic, stainless steel and silicone seal
Injection molded plastic, aluminium and thermoplastic elastomer
º
º
+5
C to +35
C
10–95% non-condensing
º
º
-20
C to +60
C
10–95% non-condensing
Flexible plastic and electrical components,1.98 m, 15 mm inner diameter
≤ 41°C
Class II (double insulation), Type BF
Please refer to the S9 Information Guide for details