Wiring the Line Power

Installation
ATTENTION EUROPEAN CUSTOMERS!
To meet CE Mark requirements, all cables must be installed
as described in Appendix A, CE Mark Compliance.
The Model GS868 may be ordered for operation with power inputs of
100-120 VAC, 220-240 VAC, or 12-28 VDC. The label on the shroud
inside the electronics enclosure, just above the
terminal block, lists the required line voltage and the fuse rating for
the unit (the fuse rating is also listed in Chapter 4, Specifications). Be
sure to connect the meter only to the specified line voltage.
Note: For compliance with the European Union's Low Voltage
Directive (73/23/EEC), this unit requires an external power
disconnect device such as a switch or circuit breaker. The
disconnect device must be marked as such, clearly visible,
directly accessible, and located within 1.8 m (6 ft) of the
Model GS868.
Refer to Figure 1-3 on page 1-8 or Figure 1-10 on page 1-20 to locate
TB1
terminal block
and connect the line power as follows:
Improper connection of the line power leads or connecting
the meter to the incorrect line voltage will damage the
unit. It will also result in hazardous voltages at the flowcell
and associated piping and within the electronics console.
TB1
line power
!WARNING!
March 2008
1-7