Checking For Intermodule Measurement Completion - HP 16500C Programmer's Manual

Logic analysis system
Hide thumbs Also See for 16500C:
Table of Contents

Advertisement

Programming Examples

Checking for Intermodule Measurement Completion

Checking for Intermodule Measurement Completion
This program can be appended to or inserted into another program when you
need to know when an intermodule measurement is complete. If it is at the
end of a program it will tell you when measurement is complete. If you insert
it into a program, it will halt the program until the current measurement is
complete.
420
! ****************** CHECK FOR MEASUREMENT COMPLETE **********************
430
! Enable the MESR register and query the register for a measurement
440
! complete condition.
450
!
460
OUTPUT 707;":SYSTEM:HEADER OFF"
470
OUTPUT 707;":SYSTEM:LONGFORM OFF"
480
!
490
Status=0
500
OUTPUT 707;":MESE0 1"
510
OUTPUT 707;":MESR0?"
520
ENTER 707;Status
530
!
533
! Start a measurement
536
OUTPUT 707;":START"
540
! Print the MESR register status.
550
!
560
CLEAR SCREEN
570
PRINT "Measurement complete status is ";Status
580
PRINT "0 = not complete, 1 = complete"
590
! Repeat the MESR query until measurement is complete.
600
WAIT 1
610
IF Status=1 THEN GOTO 630
620
GOTO 510
630
PRINT TABXY(30,15);"Measurement is complete"
640
!
650
END
15–6