General
Do not use aggressive cleaning agents that contain chlorine or solvents, nor abrasive
or sharp-edged cleaning products to clean the unit.
Always heed the product descriptions and directions for use of products you use to
desinfect.
Due to the special surface finish required for sterilizability of the BIOM 4, structural
patterns may become visible. This unavoidable minor blemish does not in any way
adversely affect the function, handling and sterilization of the unit.
Care of the BIOM 4
After cleaning and prior to sterilization, treat all moving parts of the BIOM 4 with a
sterilizable, silicone-oil-free conditioning lubricant.
Remove any excess oil from the surfaces, as staining could otherwise occur.
The following can be used for this purpose: Aesculap Sterilit i (JG600) oilspray or
Medicon Instrumentensprühöl 46.00.40.
Cleaning and sterilizing the BIOM 4
Clean the BIOM 4 with water immediately after use. This prevents incrustations
which can make sterilization impossible.
The BIOM 4 can be cleaned and disinfected in a dishwasher with a mildly alkaline
detergent, as well as in the disinfector (refer to the Conditioning manual).
All components of the BIOM 4 listed in Chapter 14 can be sterlized in a steam auto-
clave (max. 134°C / 273°F). Always remove the optical system components before
autoclaving.
Instruction manual BIOM 4 (G/54400/1011/en)
10 Care and Maintenance
25 / 34