  • Page 1 CIMEL CE318-T Quick Setup Guide...
  • Page 2 User interface: 1. Graphic display 2. Capacitive keys 3. Mechanical push button Connector panel: 4. NOT USED 5. Wet sensor 6. NOT USED 7. NOT USED 8. USB 9. Sensor head 10. Robot azimuth and zenith motion 11. DCP / Serial communication 12.
  • Page 3 GPS port (number 14) through the hole in the deployment case. 3. From the main menu scroll down to “CIMEL” and touch “enter”, then scroll down to “GPS INFO” and touch “enter” to synchronize the GPS coordinates (to turn on display, press the yellow...
  • Page 4 4. Perform a “PARK” scenario in the control box menu. The robot will find its home position 5. Strap the sensor head on the V- The 2 purple faces have to be shaped support of the robot with aligned properly the collimator pointing down.
  • Page 5 6. Perform a “PARK” scenario in the control box menu. The robot will find its home position. The collimator should return to pointing down. 7. Perform a “GOSUN” scenario in the control box menu. The sensor head will point towards the sun.
  • Page 6 8. After a “GOSUN” scenario rotate the base of the robot until the sun spot lines up to the target. Note: The sun spot may not be centered on the target. 9. ROBOT LEVELING: Fig.2 Without disturbing the sensor head, place the level Fig.3 on the AZ housing.
  • Page 7 10. Perform a “PARK” scenario in the control box menu again. The robot will find its home position. 11. Perform a “GOSUN” scenario in the control box menu. The sensor head will point to the sun.
  • Page 8 12. After a “GOSUN” scenario rotate the base of the robot until the sun spot lines up to the target. The sun spot may not be centered on the target but should be closer than 4mm. If the sun spot is further, go back to step 6. 13.
  • Page 9 14. Secure the sensor head cable to the ZN motor (fig. 1 – 3) Fig.1 Fig.2 Fig.3...
  • Page 10 15. Perform a “PARK” scenario in the control box menu again. The robot will find its home position. 16. Perform a “GOSUN” scenario in the control box menu. The sensor head will point to the sun. 17. Perform a “TRACK SUN” scenario in the Control box menu, The robot will follow the sun.
  • Page 11 19. Set the photometer in automatic mode. 20. Plug the USB adapter into the PC. Connect the PC cable to the DCP connector On the connector panel. Then connect the PC Cable to the USB adapter. 21. Run the “cimelTS_https_connect.exe” program...