Data Logging; Function Tests - MSA ALTAIR 4XR Operating Manual

Multigas detector
Hide thumbs Also See for ALTAIR 4XR:
Table of Contents

Advertisement

Bump Pairing the Device to a Bluetooth Host
This device has an integrated RFID chip to facilitate a faster Bluetooth pairing process with a Bluetooth
host that supports a RFID or NFC reader with appropriate software. Simply align the RFID or NFC
reader of the Bluetooth host directly over the approval label on the back of the device.The device and
Bluetooth host should become paired and connected.
Connecting the Device to a Bluetooth Host
If this was the last device connected to the Bluetooth host, the Bluetooth host can connect to the device
whether or not the device is in discovery mode as long as Bluetooth is enabled. The six digit code
confirmation will not be displayed.
The device will only recall the last Bluetooth host it was paired with. If connecting to
another Bluetooth host, the device must be placed into discovery mode to be detected.
Disconnecting the Device from a Bluetooth Host
The device does not have a disconnect feature as this would be initiated by the Bluetooth host. Use
the Bluetooth host functions to purposefully disconnect the device from the Bluetooth host.
Device Configuration over Bluetooth Connection
The device has the ability to receive updates to device settings over the Bluetooth connection. The
user must successfully pair the device and Bluetooth host confirming that the six digit security code
matches both on the device and the Bluetooth host. After a configuration change has been initiated,
the user must confirm the request on the device by pressing the  button.
Evacuation Alert over Bluetooth Connection
The device has the ability to receive an evacuate message over the Bluetooth connection. The user
must successfully pair the device and Bluetooth host confirming that the six digit security code matches
both on the device and the Bluetooth host. Once connected, an evacuate message sent to the device
will send the device into alarm while displaying EVAC on the display. Press the  button to silence the
Evacuation alert and confirm the alert was received. Press the  button a second time to reset the
Evacuation alert once in a safe area.
3.6

Data Logging

This device's data logs can be downloaded via the GALAXY GX2 Automated Test Stand or via a PC
using MSA Link software.
Connecting device to PC
(1)
Switch ON the device and align the Datalink Communication port on the device to the IR inter-
face of the PC.
(2)
Start the MSA Link software on the PC and start the connection by clicking the connect icon.
3.7

Function Tests

Alarm Test
Turn on the device. Verify that:
all LCD segments are activated momentarily
alarm LEDs flash
horn sounds briefly
vibrating alarm triggers briefly.
®
ALTAIR
4XR
Operation
US
33