ForaCare D20 No-Coding Owner's Manual page 50

Hide thumbs Also See for D20 No-Coding:
Table of Contents

Advertisement

► Blood Pressure Measurement
1. If nothing is displayed after pressing
POSSIBLE CAUSE
Batteries incorrectly installed or
absent.
2. If the heart rate is higher/lower than user ' s average:
POSSIBLE CAUSE
Movement during measurement.
Measurement taken just after
exercise.
3. If the result is higher/lower than user ' s average measurement:
POSSIBLE CAUSE
May not be in correct position while
measuring.
Blood pressure naturally varies from
time to time.
50
:
WHAT TO DO
Check that the batteries are correctly
installed.
WHAT TO DO
Repeat measurement.
Rest at least 30 minutes before
repeating measurement.
WHAT TO DO
Adjust to the correct position to
measure.
Keep in mind for next measurement.

Advertisement

Table of Contents
loading

Table of Contents