VC01
Model
Temperature Calibrator
Instruction
Manual
reedinstruments
www
com
REED Instruments
1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the VC01 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for REED VC01

  • Page 1 VC01 Model Temperature Calibrator Instruction Manual reedinstruments REED Instruments 1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.com...
  • Page 2: Table Of Contents

    ON/OFF ..................8 Automatic Power Off ..............8 Outputs .................8-12 Calibration ................13-15 Battery Replacement ..............16 For service (repairs or calibration) on this or any other REED product or information on other REED products, contact REED Instruments at info@reedinstruments.com reedinstruments REED Instruments...
  • Page 3: Safety

    • For optimal accuracy, allow the instrument to warm up 5 minutes before operating • If the automatic reference-junction temperature compensation of the instrument deviates from its designed accuracy, contact a qualified REED technician reedinstruments REED Instruments 1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.com...
  • Page 4: Specifications

    -10 to 50°C with its compensating error up to 0.5°C • Temperature coefficient: ±0.005% of range per °C for the temperature ranges from 0 to 18°C and from 28 to 50°C reedinstruments REED Instruments 1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.com...
  • Page 5: Specifications

    Safety: Certified compliant to IEC1010 provisions (Safety Standard issued by International Electro-technical Commission) For service on this or any other REED product or information on other REED products, contact REED Instruments at info@reedinstruments.com reedinstruments REED Instruments 1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.com...
  • Page 6: Instrument Description

    4 - RJ-ON button (TC ref-junc comp) 5 - Output Zero Reset button 6 - Function Selection button 7 - Range Selection key Output Digit Selection button Output Value Setting button 10 - Output terminal reedinstruments REED Instruments 1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.com...
  • Page 7 Indicates that the output digits need to be set Indicates the current output value (unit of measure) Indicates that the output signal is on Indicates the type of thermocouple Indicates the respective type of a RTD reedinstruments REED Instruments 1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.com...
  • Page 8: Operating Instructions

    Do not apply any voltage to the output terminal during the operation. If any improper voltage is applied to the output terminal, it will cause damage to the internal circuit. continued ... reedinstruments REED Instruments 1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.com...
  • Page 9 -200 to 1000°C 0000.0°C -200 to 1200°C 0000.0°C -200 to 400°C 400 °C 400 to 1800°C 0000.0°C -200 to 1300°C RTD Pt100 Pt100 000.0 °C -200 to 850°C Cµ50 000.0 °C -50 to 150°C reedinstruments REED Instruments 1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.com...
  • Page 10 5) Press the button to change the numerical value of the set digits. Hold the button for one second and the value will keep varying 6) Automatic compensation for reference-junction temperature continued ... reedinstruments REED Instruments 1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.com...
  • Page 11 (set resistance ) = VX (output voltage) /IX( exciting current), the calibrated device must provide an exciting current to the simulator. The exciting current should range from 0.5mA to 2mA. continued ... reedinstruments REED Instruments 1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.com...
  • Page 12 Hold the button for one second and the value will keep varying 7) Press the Zero button and the output will be directly set to 000.0°C 8) Press the °C/°F button to select the unit °C or °F reedinstruments REED Instruments 1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.com...
  • Page 13: Calibration

    24 hours. Then set the instrument to a state of non-automatic shutdown followed by warming it up for 30 minutes. NOTE Power supply for calibration: During the calibration, the battery needs replacing. reedinstruments REED Instruments 1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.com...
  • Page 14 4 & 5 NOTE Calibration Storage: Press the ZERO button to store the calibrated point when the display does not flash, denoting that the calibration storage is invalid. reedinstruments REED Instruments 1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.com...
  • Page 15: Calibration

    4 & 5 in 1V Range Calibration on Page 14 NOTE Exciting current: The flow direction of the exciting current must be in line with the calibration point, otherwise the display fails to flash, denoting that the calibration storage is invalid. reedinstruments REED Instruments 1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.com...
  • Page 16: Battery Replacement

    Notes _________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ For service on this or any other REED product or information on other REED products, contact REED Instruments at info@reedinstruments.com reedinstruments REED Instruments 1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.com...