Installation Requirements - Nano Temper Monolith NT.115 User Manual

Microscale thermophoresis instrument
Table of Contents

Advertisement

Installation Requirements

To ensure operation safety, observe the following conditions:
1.
Only operate the Monolith NT.115 instrument with the supplied external power supply (SPU63-
105, Sinpro Electronic Co Ltd or Ael60US12, XP Power LLC or GS90A12-P1M, MeanWell or
GS60A12-P1J, MeanWell).
2.
Only connect the external power supply of Monolith NT.115 to an electrical socket containing a
protective conductor terminal.
3.
Ensure that the power plug of the external power supply is easily accessible. The Monolith
NT.115 instrument has to be installed in a way that it does not hinder the access to the external
power supply and its power plug.
4.
Only operate the instrument with the delivered PC (notebook).
5.
Only operate the instrument with original Monolith NT.115 capillary trays.
6.
The maximum noise level of the instrument is 64 dB(A). Only operate the instrument in an
environment where this noise level is appropriate.
Operate the instrument in a temperature range of 15 – 30°C.
7.
Operate the instrument in a humidity range of 0 – 80 % (RH). If ambient humidity exceeds 80 %
8.
(RH), condensation may deteriorate optical components.
Operate the instrument in an atmospheric pressure range of 800 – 1060 hPa.
9.
10. Do not operate the Monolith NT.115 under conditions which pose a risk of explosion, implosion
or the release of gases.
11. Avoid strong magnetic fields and sources of high frequency. The instrument may not function
properly when near a strong magnetic field or high frequency source.
12. Avoid vibrations from vacuum pumps, centrifuges, electric motors, processing equipment and
machine tools.
13. Avoid dust and corrosive gas. Do not install the instrument where it may by exposed to dust,
especially in locations exposed to outside air or ventilation outlets.
14. To clean the instrument, only use water.
15. Do not install the instrument in a location where it may be exposed to direct sunlight.
16. Install the instrument in a horizontal and stable position. (This includes a table, bench or desk
upon which the instrument is installed).
17. Ensure that no air conditioner blows air directly onto the instrument. This may prevent stable
measurements.
18. Install the instrument in a location that allows easy access for maintenance.
Note: The above conditions do not guarantee optimal performance of this instrument.