Specifications; Input Characteristics - National Instruments USB-9211A User Manual

4-channel, 24-bit thermocouple input devices
Hide thumbs Also See for USB-9211A:
Table of Contents

Advertisement

Specifications

Input Characteristics

© National Instruments Corporation
The following specifications are typical at 25 °C, unless otherwise noted.
All voltages are relative to COM unless otherwise noted.
Number of channels ............................... 4 thermocouple channels,
ADC resolution ...................................... 24 bits
Type of ADC.......................................... Delta-sigma
Input range ............................................. ±80 mV (not software selectable)
Common-mode range
Channel-to-COM ............................ ±1.5 V
Common-to-earth ground ............... ±250 V
Common-mode rejection ratio (0 to 60 Hz)
Channel-to-common ....................... 95 dB
Common-to-earth ground ............... >170 dB
Temperature measurement ranges ......... Works over temperature
Cold-junction compensation sensor accuracy
0 to 60 °C ........................................ 0.6 °C (1.1 °F) typ,
Conversion time ..................................... 70 ms per channel;
1 internal autozero channel,
1 internal cold-junction
compensation channel
ranges defined by NIST
(J, K, R, S, T, N, E, and
B thermocouple types)
1.3 °C (2.3 °F) max
420 ms total for all channels
including the autozero and
cold-junction channels
19
NI USB-9211/9211A User Guide and Specifications

Advertisement

Table of Contents
loading

This manual is also suitable for:

Usb-9211

Table of Contents