OD600 DiluPhotometer User Manual Implen GmbH Schatzbogen 52 D-81829 Munich Declaration of Conformity This is to certify that the Implen OD600 DiluPhotometer™ conforms to the requirements of the following directives: 2014/35/EU Low Voltage directive (LVD) 2014/30/EU Electromagnetic Compatibility directive (EMC)
OD600 DiluPhotometer User Manual ESSENTIAL SAFETY NOTES Unpacking, Positioning and Installation Check the contents of the pack against the packing list. If any shortages are discovered, please inform your supplier immediately. Inspect the instrument for any signs of damage caused in transit. If any damage is discovered, please inform your supplier immediately.
OD600 DiluPhotometer User Manual INTRODUCTION Your Spectrophotometer Your OD600 is a small, easy to use instrument that is dedicated to measuring samples at a wavelength of 600 nm. It is suitable for measuring growth rates of all types of cell (e.g. E. coli and yeast). Combining OD600...
OD600 DiluPhotometer User Manual Keypad and Display Action Turns the instrument on/off On/Off key Blank Set reference to 0.000 OD at 600 nm on a reference Sample Take a measurement Memory button Reset Press twice to clear stored values Recall / Print...
DILUCELL Description DiluCell™ is especially designed for the usage of the OD600 DiluPhotometer for the analysis of samples at a wavelength range of 340 nm to 950 nm. Due to the reduced pathlength, DiluCell™ provides a virtual dilution without the need of a physical dilution of higher concentrated samples.
DC 20 Limitations Do not autoclave DiluCell™. The DiluCell™ is approved only for the usage with the OD600 DiluPhotometer™. For solvent compatibility please contact your local Implen partner or the Implen Team at www.implen.de. Version 1.3 Page 8 / 14...
OD600 DiluPhotometer User Manual OD600 - DILUCELL APPLICATION Bacterial Cell Culture Measurement 4.1.1 Description The stage of growth of a bacterial culture needs to be monitored to ensure that the cells are harvested at the optimum point for the greatest density of live cells. An exemplary growth curve is given below. Cells should be harvested towards the end of the log phase.
Page 10
OD600 DiluPhotometer User Manual Using the memory function The instrument can store up to 99 readings in the memory. The results can then be viewed, printed or downloaded at a later time. This enables readings to be taken at, for example, an incubator and downloaded to a PC in a different laboratory.
OD600 DiluPhotometer User Manual OD600 - STANDARD CUVETTE APPLICATION Protein Quantification with Bradford Assay 5.1.1 Description Bradford Assay is a rapid and accurate method used to determine the total protein concentration of a sample by quantifying the binding of a dye, Coomassie Brilliant Blue G-250, to an unknown protein. The assay is based on the observation that the absorbance maximum for an acidic solution of Coomassie Brilliant Blue G-250 shifts from 465 nm to 595 nm when binding to protein occurs.
OD600 DiluPhotometer User Manual TROUBLE SHOOTING NOTES ERROR INDICATION SOLUTION A flashing Absorbance reading of This indicates an Absorbance of more than 1.99 and is therefore out of range. 2.00 Abs is obtained. The sample needs to be diluted or use DiluCell...
OD600 DiluPhotometer User Manual ACCESSORIES DiluCell DC 10 DiluCell DC 20 Adapter set for 10 and 12 mm tubes B-80-3000-57 MAINTENANCE Cleaning and General Care of the Instrument The instrument has no serviceable parts. The instrument requires little maintenance, but the following are considered good practice: 1.
The warranty included in the conditions of supply is valid for 12 months only if the product has been used according to the instructions supplied. IMPLEN or your supplier can accept no liability for loss or damage, however caused, arising from the faulty or incorrect use of this product.